Encrypted login | home

Program Information

Clinical Use of a Real-Time Surface Image-Guided Positioning and Tracking System in Proton Therapy


W Hsi

W Hsi1*, Y Kang1, X Ding1, A Mascia1, E Ramirez1, Y Zheng1,and O Zeidan1, (1) ProCure Proton Therapy Center, Oklahoma City, Oklahoma, USA

SU-E-J-76 Sunday 3:00:00 PM - 6:00:00 PM Room: Exhibit Hall

Topic of interest: Clinical applications of AlignRT 3-cameras real-time surface image-guided positioning system (IGPS) for positioning patients to reduce the number of X-ray images and tracking intra-fractional movements in proton therapy.
Purposes: To position patients and track the intra-fractional movements, the AlignRT system was implemented in proton incline-beam-line (IBL) at Procure Oklahoma-City center.
Methods: The AlignRT3c system was configured near perpendicular to the gantry rotation for accommodating the X-ray IGPS. To evaluate positioning accuracy, more than 10 surfaces of each patient for ten patients with intracranial tumors were acquired after patients positioned by X-ray IGPS. Displacements between acquired surfaces and the reference surface taken at 1st day of treatment were examined. Intra-fractional movements with respiratory was studied with gated surface that allows setting the reference surface for patient at exhale during breathing. Intra-fractional movements due to respiratory were monitored on 10 sections of each patient for three patients with thoracic tumors.
Results: Accuracy of positioning patient is 2.0 mm at both anterior-posterior and lateral directions, and is 3.5 mm in superior-inferior (SI) direction by aligning the surfaces of masks. Observed larger displacements along SI direction can be due to patient's movements within the mask. Periodical displacements within 5 mm compared to its reference were seen for the three patients with thorax tumors. However, 10 mm sharp displacements with a few seconds were observed when patient moved the body.
Conclusions: We have implemented the first AlignRT3c IGPS for proton therapy for positioning patients within 2.0 mm, and successfully tracked intra-fractional respiratory motion during treatment after positioning patient.


Contact Email