Encrypted login | home

Program Information

Impact of Calculation Grid Size On Dosimetry and Radiobiological Parameters for Head and Neck IMRT


S Srivastava

S Srivastava1,2,4*, C Cheng1,2 , I Das1,2,4 , (1) Purdue University, West Lafayette, IN,(2) Indiana University Health Methodist Hospital, Indianapolis, IN, (3) University Hospitals Case Medical Center, Cleveland, OH, (4) Indiana University- School of Medicine, Indianapolis, IN

Presentations

SU-E-T-454 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose:
IMRT has become standard of care for complex treatments to optimize dose to target and spare normal tissues. However, the impact of calculation grid size is not widely known especially dose distribution, tumor control probability (TCP) and normal tissue complication probability (NTCP) which is investigated in this study.

Methods:
Ten head and neck IMRT patients treated with 6 MV photons were chosen for this study. Using Eclipse TPS, treatment plans were generated for different grid sizes in the range 1-5 mm for the same optimization criterion with specific dose-volume constraints. The dose volume histogram (DVH) was calculated for all IMRT plans and dosimetric data were compared. ICRU-83 dose points such as D2%, D50%, D98%, as well as the homogeneity and conformity indices (HI, CI) were calculated. In addition, TCP and NTCP were calculated from DVH data.

Results:
The PTV mean dose and TCP decreases with increasing grid size with an average decrease in mean dose by 2% and TCP by 3% respectively. Increasing grid size from 1-5 mm grid size, the average mean dose and NTCP for left parotid was increased by 6.0% and 8.0% respectively. Similar patterns were observed for other OARs such as cochlea, parotids and spinal cord. The HI increases up to 60% and CI decreases on average by 3.5% between 1 and 5 mm grid that resulted in decreased TCP and increased NTCP values. The number of points meeting the gamma criteria of ±3% dose difference and ±3mm DTA was higher with a 1 mm on average (97.2%) than with a 5 mm grid (91.3%).

Conclusion:
A smaller calculation grid provides superior dosimetry with improved TCP and reduced NTCP values. The effect is more pronounced for smaller OARs. Thus, the smallest possible grid size should be used for accurate dose calculation especially in H&N planning.


Contact Email: