Encrypted login | home

Program Information

Characterization of the PTW Synthetic Diamond Detector for Radiation Therapy Measurements


N Papanikolaou

S Stathakis , M Markovic , P Mavroidis , N Papanikolaou*, University of Texas Health Science Center, UTHSCSA, San Antonio, TX

Presentations

SU-E-T-390 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose: To investigate the dosimetric properties of new commercially available synthetic single crystal diamond detector under irradiation with therapeutic photon beams from linear accelerators.
Methods: A single crystal diamond detector was tested using 6MV photon beam. The detector performance was evaluated for reproducibility, linearity with dose, dose rate dependence, angular dependence, collection efficiency, and measurement of output factors. Lateral field profiles, and percentage depth dose profiles were measured and compared against commercially available detectors.

Results: Reproducibility of the detector measurement has a standard deviation of 0.1%. A good linear behavior of the detector response as a function of the delivered dose is observed with deviations below ±0.03% in the dose range from 0.1 to 5Gy. In addition, the detector response is dose rate independent, with deviations below 0.1% in the investigated dose rate range from 1 to 10Gy per min. Charge collection efficiency deviations were within 0.07% from 1 to 10Gy. No angular dependence along the radial direction while up to 1.3% angular dependence was observed in the axial direction. Percentage depth dose curves obtained from the diamond detector are in good agreement with the ones from the reference dosimeters. Lateral beam profile measurements show an overall good agreement among detectors, taking into account their respective geometrical features. The spatial resolution of solid state detectors is confirmed to be better than that of ionization chambers, being the one from the diamond detector comparable to that of the silicon diode.

Conclusions: Conclusions: The observed dosimetric properties indicate that the tested diamond detector is a suitable candidate for clinical photon beam dosimetry. The agreement with reference dosimeters show that the detector is suitable for measurements for large fields as well as small fields as the ones used for stereotactic radiotherapy.



Contact Email: