Encrypted login | home

Program Information

Dosimetric Characterization of Gafchromic Film EBT3 Using Vidar DosimetryPro Advantage RED and EPSON Expression 10000XL Scanners


C Venencia

L Medina , A Adrada , M Filipuzzi , E Garrigo , C Venencia*, Instituto Privado de Radioterapia - Fundacion Marie Curie, Cordoba, Argentina

Presentations

SU-E-T-121 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose: The purpose of this paper is to characterize EBT3 using two types of scanner, analyzing the factors of influence of each dosimetry system.

Methods: The film used in this study was GAFCHROMIC EBT3, the films were exposed at a dose range between 0Gy a 9Gy in a solid water phantom, SSD=100cm, 5cm depth and perpendicularly to the 6MV photon beam generated by a Novalis TX linear accelerator equipped with an HDMLC. A Farmer type ion chamber TN30013 (PTW) was used to determine the dose delivered to the film. The films were digitized with a scanner EPSON expression 10000XL and the VIDAR DosimetryPro Adventage RED. Software RIT113v6.1 was used for construction of the calibration curve and analysis. The film characteristics investigated were: response at different dose levels, sensitivity to orientation and side and resolution through the results of the spatial response function by analyzing a step pattern. Additionally, 20 IMRT treatment fields were measured with both scanner and compared with calculated dose using gamma index analysis (3%-3mm).

Results: The OD obtained for dose level 2Gy in the orientation portrait of the film on the scanner EPSON is (0,222±0,19) and for Vidar RED (0,252±0,10) and landscape is for EPSON (0,211±0,25) and for Vidar RED (0,250±0,11) . The orientation dependence with respect to film side is about 0,09% for EPSON and about 0.03% for VIDAR. The spatial response function increase in response to the Gaussian function FWHM EPSON scanner (0.18mm) compared with VIDAR scanner function (less than 0.06mm) was observed. We analyzed 20 total plan dose distributions the number of pixels with gamma>1 (3%-3mm) was 0.7%±1.2 [0.1%; 2.82%] for EBT3-VIDAR y 2%±2.9 [0.2%; 3.5%] for EBT3-EPSON.

Conclusion: VIDAR scanner shows better sensitivity. EBT3 film shows a different response between portrait and landscape orientation. Step pattern is better reproduce by VIDAR scanner.


Contact Email: