What Effect Have the Volume Defined in the Alignment Clipbox for Cervical Cancer Using Automatic Registration Methods for Cone-Beam CT Verification?

no image available
W Wang

W Wang*, H Yang , Y Wang , H Jia , Y Xie , Taizhou Hospital, Wenzhou Medical College, Taizhou, Zhejiang

Presentations

SU-E-J-119 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose:To investigate the impact of different clipbox volumes with automated registration techniques using commercially available software with on board volumetric imaging(OBI) for treatment verification in cervical cancer patients.

Methods:Fifty cervical cancer patients received daily CBCT scans(on-board imaging v1.5 system, Varian Medical Systems) during the first treatment week and weekly thereafter were included this analysis. A total of 450 CBCT scans were registered to the planning CT-scan using pelvic clipbox(clipbox-Pelvic) and around PTV clip box(clipbox-PTV). The translations(anterior-posterior, left-right, superior-inferior) and the rotations(yaw, pitch and roll) errors for each matches were recorded. The setup errors and the systematic and random errors for both of the clip-boxes were calculated. Paired Samples t test was used to analysis the differences between clipbox-Pelvic and clipbox-PTV.

Results:. The SD of systematic error(Σ) was 1.0mm, 2.0mm ,3.2mm and 1.9mm,2.3mm, 3.0mm in the AP, LR and SI directions for clipbox-Pelvic and clipbox-PTV, respectively. The average random error(σ)was 1.7mm, 2.0mm ,4.2mm and 1.7mm,3.4mm, 4.4mm in the AP, LR and SI directions for clipbox-Pelvic and clipbox-PTV, respectively. But, only the SI direction was acquired significantly differences between two image registration volumes(p=0.002,p=0.01 for mean and SD). For rotations, the yaw mean/SD and the pitch SD were acquired significantly differences between clipbox-Pelvic and clipbox-PTV.


Conclusion:The defined volume for Image registration is important for cervical cancer when 3D/3D match was used. The alignment clipbox can effect the setup errors obtained. Further analysis is need to determine the optimal defined volume to use the image registration in cervical cancer.

Funding Support, Disclosures, and Conflict of Interest: Conflict of interest: none.