Encrypted login | home

Program Information

The New Evaluation Method of the VMAT Plan Delivery Using Varian DynaLog Files and Modulation Complexity Score (MCS)

no image available
K Tateoka

K Tateoka1,2*, K Fujimomo1 , Y Saitou2 , T Nakazawa2 , T Abe2 , A Nakata2 , M Yano2 , M Hareyama1 , (1) Proton Treatment Center, Radiation Therapy Research Institute, Social Medical Corporation Teishinkai, Sapporo, JAPAN, (2) Graduate School of Medicine, Sapporo Medical University, Sapporo, JAPAN

Presentations

SU-E-T-325 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose:The aim of the study is to evaluate the use of Varian DynaLog files to verify VMAT plans delivery and modulation complexity score (MCS) of VMAT.

Methods:Delivery accuracy of machine performance was quantified by multileaf collimator (MLC) position errors, gantry angle errors and fluence delivery accuracy for volumetric modulated arc therapy (VMAT). The relationship between machine performance and plan complexity were also investigated using the modulation complexity score (MCS). Plan and Actual MLC positions, gantry angles and delivered fraction of monitor units were extracted from Varian DynaLog files. These factors were taken from the record and verify system of MLC control file. Planned and delivered beam data were compared to determine leaf position errors and gantry angle errors. Analysis was also performed on planned and actual fluence maps reconstructed from those of the DynaLog files. This analysis was performed for all treatment fractions of 5 prostate VMAT plans. The analysis of DynaLog files have been carried out by in-house programming in Visual C++.

Results:The root mean square of leaf position and gantry angle errors were about 0.12 and 0.15 , respectively. The Gamma of planned and actual fluence maps at 3%/3 mm criterion was about 99.21. The gamma of the leaf position errors were not directly related to plan complexity as determined by the MCS. Therefore, the gamma of the gantry angle errors were directly related to plan complexity as determined by the MCS.

Conclusion:This study shows Varian dynalog files for VMAT plan can be diagnosed delivery errors not possible with phantom based quality assurance. Furthermore, the MCS of VMAT plan can evaluate delivery accuracy for patients receiving of VMAT. Machine performance was found to be directly related to plan complexity but this is not the dominant determinant of delivery accuracy.


Contact Email: