Encrypted login | home

Program Information

Analysis of Robustness of Proton Pencil Beam Scanning Technique for Delivery of Craniospinal Irradiation

no image available
H Lin

H Lin*, M Kirk , H Zhai , X Ding , H Liu , C Hill-Kayser , R Lustig , Z Tochner , J McDonough , S Both , Univ Pennsylvania, Philadelphia, PA

Presentations

SU-E-T-621 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose: To investigate the robustness and safety of craniospinal irradiation (CSI) planned with a proton pencil beam scanning (PBS) technique which overcomes the complexity of the planning associated with feathering match lines.

Methods: Six CSI patients were planned with gradient-dose matching using PBS technique. Uniform dose coverage to the entire target volumes was achieved with averaged junction lengths of 6.9±0.3 cm. Robustness of the plans was evaluated by shifting the isocenter of each treatment field by ±3 mm in longitudinal direction and compared with the original non-shifted plan with metrics of conformity number (CN) and homogeneity index (HI). An anthropomorphic phantom study using film measurements was also carried out on a plan with 5 cm junction length.

Results: For a given junction length, the dose errors were directly proportional to the setup errors. Setup errors of 3 mm from each field caused on average 3.5% lower CN and 2.1% higher HI. Minimal D95% to PTV and D98% to CTV were reduced by 2.2%±1.5% and 2.8%±1.7% respectively. A drop of maximal 6.8%±5.5% on the minimal dose to the cribriform plate was also observed. When the junction length was 5cm or longer, these 3mm setup errors from each field resulted in up to 12% dose errors. Consistent results were reached between film measurements and planned dose profiles in the junction area. Due to near-zero exit doses beyond the target volume, sparing of anterior organs such as heart, liver, lung and kidney were observed.

Conclusions: Longitudinal setup errors directly reduce the dosimetric accuracy of the CSI treatment with matched proton fields. The reported technique creates a slow dose gradient in the junction area, which makes the treatment more robust and safe to longitudinal setup errors compared to conventional feathering methods.



Contact Email: