Encrypted login | home

Program Information

CTDI Values for All Protocols: Using the Ratio of the DLP Measured in CTDI Phantoms to the Measured Air Exposure

no image available
G Raterman

G Raterman*, D Gauntt , UAB Medical Center, Birmingham, AL

Presentations

SU-E-I-19 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose: To propose a method other than CTDI phantom measurements for routine CT dosimetry QA. This consists of taking a series of air exposure measurements and calculating a factor for converting from this exposure measurement to the protocol’s associated head or body CTDI value using DLP. The data presented are the ratios of phantom DLP to air exposure ratios for different scanners, as well as error in the displayed CTDI.

Methods: For each scanner, the CTDI is measured at all available tube voltages using both the head and body phantoms. Then, the exposure is measured using a pencil chamber in air at isocenter. A ratio of phantom DLP to exposure in air for a given protocol may be calculated and used for converting a simple air dose measurement to a head or body CTDI value. For our routine QA, the exposure in air for different collimations, mAs, and kVp is measured, and displayed CTDI is recorded. Therefore, the ratio calculated may convert these exposures to CTDI values that may then be compared to the displayed CTDI for a large range of acquisition parameter combinations.

Results: It was found that all scanners tend to have a ratio factor that slightly increases with kVp. Also, Philips scanners appear to have less of a dependence on kVp; whereas, GE scanners have a lower ratio at lower kVp. The use of air exposure times the DLP conversion yielded CTDI values that were less than 10% different from the displayed CTDI on several scanners.

Conclusion: This method may be used as a primary method for CT dosimetry QA. As a result of the ease of measurement, a dosimetry metric specific to that scanner may be calculated for a wide variety of CT protocols, which could also be used to monitor display CTDI value accuracy.



Contact Email: