Evaluation of the Relative Output Ratio for Collimator Jaw and MLC Defined Small Static 6MV Photon Fields

no image available
G Cho

G Cho1,2*, D Thwaites2 , (1) Chris O'Brien Lifehouse at RPAH, Camperdown, NSW 2050, (2) Institute of Medical Physics, School of Physics, University of Sydney, Sydney, NSW 2006, Australia

Presentations

SU-E-T-175 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose:To evaluate relative output ratio of collimator jaw and MLC defined small photon fields.

Methods:Relative output ratios were measured using Gafchromic EBT3 film for a 6 MV photon beam on a Novalis Tx with HD120 MLC. Beam collimation was achieved by the jaws for 1.0 cm and 3.0 cm and MLC defined square field sizes between 0.5 cm and 1.0 cm with varying jaw settings between 2.0 and 4.0 cm. Film pieces were exposed to 4 Gy. Experiments were repeated with each session consisting of five consecutive exposures for the given MLC and/or jaw collimation and with the MLC and the jaws reset for each exposure. Films were scanned using EPSON 10000XL flatbed scanner approximately 24 hours after exposure in 48 bit RGB format at 150 dpi. Film calibration data were corrected for daily linac output variations. Doses were evaluated using the green channel with square ROI sizes of 0.1 – 0.6 cm. Converted doses were normalised for output ratio calculation using the 3.0 cm field as a machine specific reference field size. Mean output ratio and coefficient of variation (CV) were calculated for each experimental session.

Results:For the Novalis 6 MV photon beam the output ratios between 0.719 and 0.872 have been measured for the jaw/MLC combinations tested. For a jaw setting of 4.0 cm field, the mean CV of the output ratios increased from 0.77% to 1.48% with decreasing MLC field size from 1.0 cm to 0.5 cm. For a nominal MLC 1.0 cm field, the CV increased to 1.00% from 0.77% with reducing jaw field size from 4.0 cm to 2.0 cm.

Conclusion:The relative output ratio and the associated CV were dependent on the collimator jaw and MLC settings. The field size dependent CV showed similar trends to those reported in the literature.