A Method of Maximize the Noncoplanar Beam Orientations and Assure the Beam Delivery Clearance for Stereotactic Body Radiation Therapy (SBRT)

no image available
J Zhu

J Zhu*, Presence St. Joseph Medical Ctr., Joliet, IL

Presentations

SU-F-P-28 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:
Develop a method to maximize the noncoplanar beam orientations and assure the beam delivery clearance for SBRT, therefore, optimize the dose conformality to the target, increase the dose sparing to the critical normal organs and reduce the hot spots in the body.
Methods:
A SBRT body frame (Elekta, Stockholm, Sweden) was used for patient immobilization and target localization. The SBRT body frame has CT fiducials on its side frames. After patient’s CT scan, the radiation treatment isocenter was defined and its coordinators referring to the body frame was calculated in the radiation treatment planning process. Meanwhile, initial beam orientations were designed based on the patient target and critical organ anatomy. The body frame was put on the linear accelerator couch and positioned to the calculated isocenter. Initially designed beam orientations were manually measured by tuning the body frame position on the couch, the gantry and couch angles. The finalized beam orientations were put into the treatment planning for dosimetric calculations.
Results:
Without patient presence, an optimal set of beam orientations were designed and validated. The radiation treatment plan was optimized and guaranteed for delivery clearance.
Conclusion:
The developed method is beneficial and effective in SBRT treatment planning for individual patient. It first allows maximizing the achievable noncoplanar beam orientation space, therefore, optimize the treatment plan for specific patient. It eliminates the risk that a plan needs to be modified due to the gantry and couch collision during patient setup.