Sensitivity and Stability of OSLDs with Filled Deep Electron/hole Traps Under Pre-Irradiation and Bleaching Conditions
J Kim1*, S Park1 , H Lee2 , H Kim2 , C Choi1 , J Park1 , (1) Seoul National University Hospital, Seoul (2) Seoul National University Hospital, Seoul
Presentations
SU-F-T-318 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall
PURPOSE: This work evaluated the characteristics of optically stimulated luminescence dosimeters (OSLDs) with fully filled deep electron/hole traps (OSLDfull) with the bleaching conditions according to the accumulated dose.
METHODS: The OSLDs were first pre-irradiated with a Co-60 gamma ray at more than 5 kGy, so as to fill the deep electron and hole traps. Using a 6-MV beam, the OSLDfull characteristics were investigated in terms of the full bleaching, fading, dose linearity, and dose sensitivity obtained in response to the accumulated dose values. To facilitate a comparison of the dose sensitivity, OSLDs with un-filled deep electron/hole traps (OSLDempty) were investigated in the same manner. A long-pass filter was used to exclude bleaching-source wavelengths of less than 520 nm. Various bleaching time and wavelength combinations were used in order to determine the optimal bleaching conditions for the OSLDfull.
RESULTS: The fading for the OSLDfull exhibited stable signals after 8 min, for both 1- and 10-Gy. For 4-h bleaching time and an unfiltered bleaching device, the supra-linear index values for the OSLDfull were 1.003, 1.002, 0.999, and 1.001 for doses of 2, 4, 7, and 10 Gy, respectively. For a 65-Gy accumulated dose with a 5-Gy fraction, no variation in dose sensitivity was obtained for the OSLDfull, within a standard deviation of 0.85%, whereas the OSLDempty dose sensitivity decreased by approximately 2.3% per 10 Gy. The filtered bleaching device yielded a highly stable sensitivity for OSLDfull, independent of bleaching time and within a standard deviation of 0.71%, whereas the OSLDempty dose sensitivity decreased by approximately 4.2% per 10 Gy for an accumulated dose of 25 Gy with a 5-Gy fraction.
CONCLUSIONS: Under the bleaching conditions determined in this study, clinical dosimetry with OSLDfull is highly stable, having an accuracy of 1% with no change in dose sensitivity or linearity at clinical doses.
Funding Support, Disclosures, and Conflict of Interest: This work was supported by a National Research Foundation of Korea (NRF) grant, funded by the Korea government (MISP) (No. 2014M2B2A4031164), and by the Korea Health Technology R&D Project through the Korea Health Industry Development Institute (KHIDI), funded by the Ministry of Health & Welfare, Republic of Korea (HI14C3459).