MRI Positive Contrast Visualization of Prostate Brachytherapy Seeds Using An Integrated Laplacian-Based Phase Processing
A Soliman1,2*, R Nosrati2,3, H Safigholi1,2 , A Owrangi2,4 , G Morton2,4 , W Song1-4 , (1) Sunnybrook Research Institute, Toronto, ON, (2) Sunnybrook Health Sciences Center, Toronto, ON, (3) Ryerson University, Toronto, ON, (4) University of Toronto, Toronto, ON
Presentations
SU-F-I-19 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall
Purpose: To propose a new method that provides a positive contrast visualization of the prostate brachytherapy seeds using the phase information from MR images. Additionally, the feasibility of using the processed phase information to distinguish seeds from calcifications is explored.
Methods: A gel phantom was constructed using 2% agar dissolved in 1 L of distilled water. Contrast agents were added to adjust the relaxation times. Four iodine-125 (Eckert & Ziegler SML86999) dummy seeds were placed at different orientations with respect to the main magnetic field (B0). Calcifications were obtained from a sheep femur cortical bone due to its close similarity to human bone tissue composition. Five samples of calcifications were shaped into different dimensions with lengths ranging between 1.2 – 6.1 mm.
MR imaging was performed on a 3T Philips Achieva using an 8-channel head coil. Eight images were acquired at eight echo-times using a multi-gradient echo sequence. Spatial resolution was 0.7 x 0.7 x 2 mm, TR/TE/dTE = 20.0/2.3/2.3 ms and BW = 541 Hz/pixel. Complex images were acquired and fed into a two-step processing pipeline: the first includes phase unwrapping and background phase removal using Laplacian operator (Wei et al. 2013). The second step applies a specific phase mask on the resulting tissue phase from the first step to provide the desired positive contrast of the seeds and to, potentially, differentiate them from the calcifications.
Results: The phase-processing was performed in less than 30 seconds. The proposed method has successfully resulted in a positive contrast of the brachytherapy seeds. Additionally, the final processed phase image showed difference between the appearance of seeds and calcifications. However, the shape of the seeds was slightly distorted compared to the original dimensions.
Conclusion: It is feasible to provide a positive contrast of the seeds from MR images using Laplacian operator-based phase processing.