HybridArc Planning Criteria for Brain SRS

no image available
K Jeong

K Jeong*, A Basavatia , D Mynampati , M Garg , S Kalnicki , W Tome , Montefiore Medical Center, Bronx, NY

Presentations

SU-F-T-589 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:
To compare VMAT SRS plans, dynamic conformal arc (DCA) plans, and Brainlab iPlan’s capability of planning and delivering brain SRS plans by employing HybridArc. HybridArc utilizes both DCA and IMRT. Using HybridArc, the amount of DCA versus IMRT needs to be optimized.

Methods:
Four SRS patients with the aim of reducing brainstem dose were selected for this study. All patients were contoured in iPlan and transferred to Eclipse for VMAT planning. In iPlan, DCA plans were created for each case. Moreover, nine HybridArc plans with DCA-IMRT ratios between 9:1 through 1:9 were created with a single ring structure generated by subtracting 3 mm expansion of target from a 10 mm expansion of the target. Two static IMRT beams were used in each of the five DCA arcs for HybridArc. The dose was prescribed to DCA only and HybridArc plans and normalized so that the target volume (TV) receives 100% dose to 99.5% of the TV to achieve 120% ~ 130% max dose within targets. Following metrics were compared: PITV, V12Gy, CGIc, CGIg, CGI, brainstem max dose, and total monitor units (MUs).

Results:
A brainstem max dose comparable with VMAT from 30% IMRT and less with 50% or more IMRT could be achieved. PITV decreased with increasing IMRT portion and begins to saturate past an IMRT portion of 30%. The CGIg index, which represents how fast the dose falls off, was better with HybridArc in all HybridArc plans. Total MUs increased with increasing IMRT but less than VMAT in all cases.

Conclusion:
Overall, a lower brainstem max dose and a lower V12Gy with fewer MUs can be achieved with HybridArc. Considering all factors, it would be best to use a DCA-IMRT ratio of either 7:3 or 6:4.