Investigation of Gafchromic EBT3 Film Energy Dependence Using Proton, Photon, and Electron Beams


C Ferreira

C Ferreira*, E Schnell, S Ahmad, T De La Fuente Herman, University of Oklahoma Health Science Center, Oklahoma City, OK

Presentations

SU-F-I-70 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: To investigate the energy dependence of Gafchromic EBT3 film over a range of clinically used proton, photon and electron energies.

Methods: Proton beam energies of 117 and 204 MeV, corresponding respectively to ranges in water of 10 cm and 27 cm from a Mevion S250 double scatter system unit were used. Electron energies of 6 and 20 MeV and photon energies of 6 and 18 MV from a Varian Clinac 21EX Linac were used. Two pieces of film (5x5 cm²) were irradiated sequentially for doses of 100, 500, and 1000 cGy for all energies and modalities. Films were placed on the central beam axis for a 10x10 cm² field size in the middle of spread out Bragg peak (SOBP) for proton and in respective dmax for photon and electron energies. Films were scanned on a flatbed Epson Expression 10000 XL scanner on the central region of the scanning window using 48-bit, 300 dpi, and landscape orientation after 48 hours post-irradiation of film to account for optical density (OD) stabilization. Film analysis of the red channel was performed using ImageJ 1.48v (National Institutes of Health).

Results: The energy dependency of EBT3 among all energies and modalities for all doses studied was small within measurement uncertainties (1σ = ± 4.1%). The mean net OD in red channel for films receiving the same dose in the same energy modality had standard deviations within 0.9% for photons, 4.9% for electrons and 1.8% for protons. It was observed that film pieces were activated during proton irradiation, e.g., 7 mR/hr at surface after 30 minutes of irradiation, lasting for 2 hours post irradiation.

Conclusion: EBT3 energy dependency was evaluated for clinically used proton, photon, and electron energies. The film self-activation may have contributed to fog and negligible dose.