A Mathematical Model of Linac Jaw Calibration Integrated with Collimator Walkout
Y Zhao*, R Corns, V Huang, Fraser Valley Cancer Centre - BC Cancer Agency, Surrey, BC
Presentations
SU-F-E-20 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall
Purpose: Accurate jaw calibration is possible, but it does not necessarily achieve good junctions because of collimator rotation walkout. We developed a mathematical model seeking to pick an origin for calibration that minimizes the collimator walkout effect.
Methods: We use radio-opaque markers aligned with crosshair on the EPID to determine the collimator walkout at collimator angles 0°, 90° and 270°. We can accurately calibrate jaws to any arbitrary origin near the radiation field centre. While the absolute position of an origin moves with the collimator walkout, its relative location to the crosshair is an invariant. We studied two approaches to select an optimal origin. One approach seeks to bring all three origin locations (0°-90°-270°) as close as possible by minimizing the perimeter of the triangle formed by these points. The other approach focuses on the gap for 0°-90° junctions.
Results: Our perimeter cost function has two variables and non-linear behaviour. Generally it does not have zero-perimeter-length solution which leads to perfect jaw matches. The zero solution can only be achieved, if the collimator rotates about a single fixed axis. In the second approach, we can always get perfect 0°-0° and 0°-90° junctions, because we ignore the 0°-270° situation. For our TrueBeams, both techniques for selecting an origin improved junction dose inhomogeneities to less than ±6%.
Conclusion: Our model considers the general jaw matching with collimator rotations and proposes two potential solutions. One solution optimizes the junction gaps by considering all three collimator angles while the other only considers 0°-90°. The first solution will not give perfect matching, but can be clinically acceptable with minimized collimator walkout effect, while the second can have perfect junctions at the expense of the 0°-270° junctions. Different clinics might choose between these two methods basing on their clinical practices.