Optimized Design of a Diamond Detector Specifically Dedicated to the Dose Distribution Measurements in Clinical Proton Pencil Beams

no image available
C Moignier

C Moignier1*, D TROMSON1 , F MARSOLAT2 , L De Marzi2 , M POMORSKI1 , M AGELOU1 , J GARCIA HERNANDEZ1 , D LAZARO1 , A MAZAL2 , (1) Institut CEA LIST, Gif-sur-Yvette, FRANCE, (2) Institut Curie - Centre de Protontherapie d'Orsay, Orsay, FRANCE

Presentations

SU-F-T-178 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:
In proton-therapy, pencil beam scanning (PBS) dosimetry presents a real challenge due to the small size of the beam (about 3 to 8 mm in FWHM), the pulsed high dose rate (up to 100 Gy/s) and the proton energy variation (about 30 MeV to 250 MeV). In the framework of French INSERM DEDIPRO project, a specifically dedicated single crystal diamond dosimeter (SCDDo) was developed with the objective of obtaining accurate measurements of the dose distribution in PBS modality.

Methods:
Monte Carlo simulations with MCNPX were performed. A small proton beam of 5 mm in FWHM was simulated as well as diamond devices with various size, thickness and holder composition. The calculated doses-to-diamond were compared with the doses-to-water in order to reduce the perturbation effects. Monte-Carlo simulations lead to an optimized SCDDo design for small proton beams dosimetry.
Following the optimized design, SCDDos were mounted in water-equivalent holders with electrical connection adapted to standard electrometer. First, SCDDos performances (stability, repeatability, signal-to-background ratio...) were evaluated with conventional photon beams. Then, characterizations (dose linearity, dose rate dependence...) with wide proton beams were performed at proton-therapy center (IC-CPO) from Curie Institute (France) with the passive proton delivery technique, in order to confirm dosimetric requirements.
Finally, depth-dose distributions were measured in a water tank, for native and modulated Bragg Peaks with the collimator of 12 cm, and compared to a commercial PPC05 parallel-plate ionization chamber reference detector. Lateral-dose profiles were also measured with the collimator of 5 mm, and compared to a commercial SFD diode.

Results:
The results show that SCDDo design does not disturb the dose distributions.

Conclusion:
The experimental dose distributions with the SCDDo are in good agreement with the commercial detectors and no energy dependence was observed with this device configuration.