The Feasibility Research of Checking Cervical Cancer IMRT Pre-Treatment Dose Verification by Automated Treatment Planning Verification System

no image available
x Liu

x liu*, Y Yin , x lin , Shandong Cancer Hospital and Institute, China, Jinan, Shandong

Presentations

SU-F-T-440 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:To assess the preliminary feasibility of automated treatment planning verification system in cervical cancer IMRT pre-treatment dose verification.

Methods:The study selected randomly clinical IMRT treatment planning data for twenty patients with cervical cancer, all IMRT plans were divided into 7 fields to meet the dosimetric goals using a commercial treatment planning system(Pianncle ,Version 9.2and the Eclipse ,Version 13.5) .The plans were exported to the Mobius 3D (M3D)server , percentage differences of volume of a region of interest (ROI) and dose calculation of target region and organ at risk were evaluated, in order to validate the accuracy automated treatment planning verification system.

Results:The difference of volume for Pinnacle to M3D was less than results for Eclipse to M3D in ROI,the biggest difference was 0.22± 0.69%, 3.5±1.89% for Pinnacle and Eclipse respectively.M3D showed slightly better agreement in dose of target and organ at risk compared with TPS.But after recalculating plans by M3D, dose difference for Pinnacle was less than Eclipse on average,results were within 3% .

Conclusion:The method of utilizing the automated treatment planning system to validate the accuracy of plans is convenient ,but the scope of differences still need more clinical patient cases to determine.At present, it should be used as a secondary check tool to improve safety in the clinical treatment planning .