Dosimetric Comparison of Point Based Vs. Target Based Prescription for Intracavitary Brachytherapy in Cancer of the Cervix


M Ashenafi

M Ashenafi*, D McDonald , J Peng , C Mart , N Koch , L Cooper , K Vanek , Medical University of South Carolina, Charleston, SC

Presentations

SU-F-T-36 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:Improved patient imaging used for planning the treatment of cervical cancer with Tandem and Ovoid (T&O) Intracavitary high-dose-rate brachytherapy (HDR) now allows for 3D delineation of target volumes and organs-at-risk. However, historical data relies on the conventional point A-based planning technique. A comparative dosimetric study was performed by generating both target-based (TBP) and point-based (PBP) plans for ten clinical patients.

Methods:Treatment plans created using Elekta Oncentra v. 4.3 for ten consecutive cervical cancer patients were analyzed. All patients were treated with HDR using the Utrecht T&O applicator. Both CT and MRI imaging modalities were utilized to delineate clinical target volume (CTV) and organs-at-risk (rectum, sigmoid, bladder, and small bowel). Point A (left and right), vaginal mucosa, and ICRU rectum and bladder points were defined on CT. Two plans were generated for each patient using two prescription methods (PBP and TBP). 7Gy was prescribed to each point A for each PBP plan and to the target D90% for each TBP plan. Target V90%, V100%, and V200% were evaluated. In addition, D0.1cc and D2cc were analyzed for each organ-at-risk. Differences were assessed for statistical significance (p<0.05) by use of Student’s t-test.

Results:Target coverage was comparable for both planning methods, with each method providing adequate target coverage. TBP showed lower absolute dose to the target volume than PBP (D90% = 7.0Gy vs. 7.4Gy, p=0.028), (V200% = 10.9cc vs. 12.8cc, p=0.014), (ALeft = 6.4Gy vs. 7Gy, p=0.009), and (ARight = 6.4Gy vs. 7Gy, p=0.013). TBP also showed a statistically significant reduction in bladder, rectum, small bowel, and sigmoid doses compared to PBP. There was no statistically significant difference in vaginal mucosa or ICRU-defined rectum and bladder dose.

Conclusion:Target based prescription resulted in substantially lower dose to delineated organs-at-risk compared to point based prescription, while maintaining similar target coverage.