An Efficient Atlas-Based Synthetic CT Generation Method


X Han

X Han, Elekta Inc., Maryland Heights, MO

Presentations

TU-AB-BRA-2 (Tuesday, August 2, 2016) 7:30 AM - 9:30 AM Room: Ballroom A


Purpose: A major obstacle for MR-only radiotherapy is the need to generate an accurate synthetic CT (sCT) from MR image(s) of a patient for the purposes of dose calculation and DRR generation. We propose here an accurate and efficient atlas-based sCT generation method, which has a computation speed largely independent of the number of atlases used.

Methods: Atlas-based sCT generation requires a set of atlases with co-registered CT and MR images. Unlike existing methods that align each atlas to the new patient independently, we first create an average atlas and pre-align every atlas to the average atlas space. When a new patient arrives, we compute only one deformable image registration to align the patient MR image to the average atlas, which indirectly aligns the patient to all pre-aligned atlases. A patch-based non-local weighted fusion is performed in the average atlas space to generate the sCT for the patient, which is then warped back to the original patient space. We further adapt a PatchMatch algorithm that can quickly find top matches between patches of the patient image and all atlas images, which makes the patch fusion step also independent of the number of atlases used.

Results: Nineteen brain tumour patients with both CT and T1-weighted MR images are used as testing data and a leave-one-out validation is performed. Each sCT generated is compared against the original CT image of the same patient on a voxel-by-voxel basis. The proposed method produces a mean absolute error (MAE) of 98.6±26.9 HU overall. The accuracy is comparable with a conventional implementation scheme, but the computation time is reduced from over an hour to four minutes.

Conclusion: An average atlas space patch fusion approach can produce highly accurate sCT estimations very efficiently. Further validation on dose computation accuracy and using a larger patient cohort is warranted.