Assessment of Inter- and Intra-Fractional Motion for Extremity Soft Tissue Sarcoma Patients by Using In-House Real-Time Optical Image-Based Monitoring System


H Kim

H Kim1*, I Kim2 , S Ye2,3 , (1) Interdisciplinary Program in Radiation Applied Life Science, College of Medicine, Seoul National University, Seoul, Republic of Korea, (2) Dept. of Radiation Oncology, Seoul National University Hospital, Seoul, Republic of Korea, (3) Program in Biomedical Radiation Sciences, Graduate School of Convergence Science and Technology, Seoul National University, Seoul, Republic of Korea

Presentations

SU-G-JeP4-14 (Sunday, July 31, 2016) 5:30 PM - 6:00 PM Room: ePoster Theater


Purpose: This study aimed to assess inter- and intra-fractional motion for extremity Soft Tissue Sarcoma (STS) patients, by using in-house real-time optical image-based monitoring system (ROIMS) with infra-red (IR) external markers.

Methods: Inter- and intra-fractional motions for five extremity (1 upper, 4 lower) STS patients received postoperative 3D conformal radiotherapy (3D-CRT) were measured by registering the image acquired by ROIMS with the planning CT image (REG_ROIMS). To compare with the X-ray image-based monitoring, pre- and post-treatment cone beam computed tomography (CBCT) scans were performed once per week and registered with planning CT image as well (REG_CBCT). If the CBCT scan is not feasible due to the large couch shift, AP and LR on-board imager (OBI) images were acquired. The comparison was done by calculating mutual information (MI) of those registered images.

Results: The standard deviation (SD) of the inter-fractional motion was 2.6 mm LR, 2.8 mm SI, and 2.0 mm AP, and the SD of the intra-fractional motion was 1.4 mm, 2.1 mm, and 1.3 mm in each axis, respectively. The SD of rotational inter-fractional motion was 0.6° pitch, 0.9° yaw, and 0.8° roll and the SD of rotational intra-fractional motion was 0.4° pitch, 0.9° yaw, and 0.7° roll. The derived averaged MI values were 0.83, 0.92 for REG_CBCT without rotation and REG_ROIMS with rotation, respectively.

Conclusion: The in-house real-time optical image-based monitoring system was implemented clinically and confirmed the feasibility to assess inter- and intra-fractional motion for extremity STS patients while the daily basis and real-time CBCT scan is not feasible in clinic.