Extrapolation Techniques for Small Field Dosimetry Using Gafchromic EBT3 Film

no image available
J Morales

J Morales1*, (1) Chris OBrien Lifehouse, Camperdown, NSW

Presentations

SU-F-T-579 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:
The purpose of this project is to test an experimental approach using an extrapolation technique for Gafchromic EBT3 film for small field x-ray dosimetry.

Methods:
Small fields from a Novalis Tx linear accelerator with HD Multileaf Collimators with 6 MV was used. The field sizes ranged from 5 x 5 to 50 x 50 mm2 MLC fields and a range of circular cones of 4 to 30 mm2 diameters. All measurements were performed in water at an SSD of 100 cm and at a depth of 10 cm. The relative output factors (ROFs) were determined from an extrapolation technique developed to eliminate the effects of partial volume averaging in film scan by scanning films with high resolution (1200 DPI). The size of the regions of interest (ROI) was varied to produce a plot of ROFs versus ROI which was then extrapolated to zero ROI to determine the relative output factor. The results were compared with other solid state detectors with proper correction, namely, IBA SFD diode, PTW 60008 and PTW 60012 diode.

Results:
For the 4 mm cone, the extrapolated ROF had a value of 0.658 ± 0.014 as compared to 0.642 and 0.636 for 0.5 mm and 1 mm2 ROI analysis, respectively. This showed a change in output factor of 2.4% and 3.3% at this comparative ROI sizes. In comparison, the 25 mm cone had a difference in measured output factor of 0.3% and 0.5% between 0.5 and 1.0 mm, respectively compared to zero volume. For the fields defined by MLCs a difference of up to 2% for 5x5 mm2 was observed.

Conclusion:
A measureable difference can be seen in ROF based on the ROI when radiochromic film is used. Using extrapolation technique from high resolution scanning a good agreement can be achieved.