An Evaluation of a Low Dose Rate (LDR) Prostate Brachytherapy Procedure Using a Failure Modes and Effects Analysis (FMEA)
S-K Cheong*, J Kim , University of Pittsburgh Medical Center, Pittsburgh, PA
Presentations
SU-G-TeP4-5 (Sunday, July 31, 2016) 5:30 PM - 6:00 PM Room: ePoster Theater
Purpose: The aim of the study is the application of a Failure Modes and Effects Analysis (FMEA) to access the risks for patients undergoing a Low Dose Rate (LDR) Prostate Brachytherapy Treatment.
Methods: FMEA was applied to identify all the sub processes involved in the stages of identifying patient, source handling, treatment preparation, treatment delivery, and post treatment. These processes characterize the radiation treatment associated with LDR Prostate Brachytherapy. The potential failure modes together with their causes and effects were identified and ranked in order of their importance. Three indexes were assigned for each failure mode: the occurrence rating (O), the severity rating (S), and the detection rating (D). A ten-point scale was used to score each category, ten being the number indicating most severe, most frequent, and least detectable failure mode, respectively. The risk probability number (RPN) was calculated as a product of the three attributes: RPN = O X S x D. The analysis was carried out by a working group (WG) at UPMC.
Results: The total of 56 failure modes were identified including 32 modes before the treatment, 13 modes during the treatment, and 11 modes after the treatment. In addition to the protocols already adopted in the clinical practice, the prioritized risk management will be implanted to the high risk procedures on the basis of RPN score.
Conclusion: The effectiveness of the FMEA method was established. The FMEA methodology provides a structured and detailed assessment method for the risk analysis of the LDR Prostate Brachytherapy Procedure and can be applied to other radiation treatment modes.