Clinical Experience with Radiation Dose Reduction of CT Examinations Using Iterative Reconstruction Algorithms

no image available
V Weir

V Weir1*, J Zhang2 , (1) Baylor Scott and White Healthcare System, Dallas, TX, (2) University of Kentucky, Lexington, KY

Presentations

SU-F-P-45 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:
Iterative reconstruction (IR) algorithms have been adopted by medical centers in the past several years. IR has a potential to substantially reduce patient dose while maintaining or improving image quality. This study characterizes dose reductions in clinical settings for CT examinations using IR.

Methods:
We retrospectively analyzed dose information from patients who underwent abdomen/pelvis CT examinations with and without contrast media in multiple locations of our Healthcare system. A total of 743 patients scanned with ASIR on 64 slice GE lightspeed VCTs at three sites, and 30 patients scanned with SAFIRE on a Siemens 128 slice Definition Flash in one site was retrieved. For comparison, patient data (n=291) from a GE scanner and patient data (n=61) from two Siemens scanners where filtered back-projection (FBP) was used was collected retrospectively. 30% and 10% ASIR, and SAFIRE Level 2 was used. CTDIvol, Dose-length-product (DLP), weight and height from all patients was recorded. Body mass index (BMI) was calculated accordingly. To convert CTDIvol to SSDE, AP and lateral dimensions at the mid-liver level was measured for each patient.

Results:
Compared with FBP, 30% ASIR reduces dose by 44.1% (SSDE: 12.19mGy vs. 21.83mGy), while 10% ASIR reduced dose by 20.6% (SSDE 17.32mGy vs. 21.83). Use of SAFIRE reduced dose by 61.4% (SSDE: 8.77mGy vs. 22.7mGy). The geometric mean for patients scanned with ASIR was larger than for patients scanned with FBP (geometric mean is 297.48 mmm vs. 284.76 mm). The same trend was observed for the Siemens scanner where SAFIRE was used (geometric mean: 316 mm with SAFIRE vs. 239 mm with FBP). Patient size differences suggest that further dose reduction is possible.

Conclusion:
Our data confirmed that in clinical practice IR can significantly reduce dose to patients who undergo CT examinations, while meeting diagnostic requirements for image quality.