Dosimetric Evaluation of Intracranial Stereotactic Radiotherapy Plans On a LINAC


N Sheth

N Sheth*, A Tabibian , J Rose , M Alvelo , C Perel , K Laiken , A Kim , Bayonne Medical Center, Bayonne, New Jersey

Presentations

SU-F-T-604 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: To evaluate the dosimetry of cranial stereotactic radiotherapy (SRT) plans of varying techniques on linac that meets appropriate TG-142 tolerances using 1 cm leaf width multileaf collimator (MLC).

Methods: Seventeen spherical targets were generated in the center of a head phantom with diameters ranging 8 mm to 40 mm. SRT plans used 100° non-coplanar arcs and 5 couch angles with 35° spacing. The field size was target plus 1 mm margin. Four plans were created for each target: symmetrical jaws blocking for 5 arcs with 0° collimator (J1C), symmetrical jaws blocking with 5 clockwise arcs with 0° collimator and 5 counter-clockwise arcs with 45° collimator (J2C), MLC blocking for 5 dynamic conformal arcs with 0° collimator (M1C), and MLC blocking for 5 clockwise dynamic conformal arcs with 0° collimators and 5 counter-clockwise dynamic conformal arcs with 45° collimator (M2C).

Conformity was evaluated using a ratio of Rx to target volume (PITV). Heterogeneity was determined using a ratio of maximum dose to Rx dose. Falloff was scored using CGIg: difference of effective radii of spheres equal to half and full Rx volumes.

Results: All plans met RTOG SRS criteria for conformity and heterogeneity. The mean PITV was 1.52±0.07, 1.49±0.08, 1.39±0.05, and 1.37±0.04 for J1C, J2C, M1C, and M2C plans respectively. The mean CGIg was 75.35±15.79, 74.19±16.66, 77.14±15.12, and 76.28±15.78 for J1C, J2C, M1C, and M2C plans respectively. The mean MDPD was 1.25±0.00 for all techniques.

Conclusion: Clinically acceptable SRT plans for spherical targets were created on a linac with 1 cm MLC. Adding two collimator angles and MLC to arcs each improved conformity. The MLC improved the dose falloff while two collimator angles degraded it. This technique can expand the availability of SRT to patients especially to those who cannot travel to a facility with a dedicated stereotactic radiosurgery machine.