Measurement of Proton Beam Spot Size Using a 2D Ion Chamber Array for Daily QA

no image available
N Thorne

N Thorne*, A Kassaee , University Pennsylvania, Philadelphia, PA

Presentations

SU-F-T-167 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: To develop an algorithm which can calculate the Full Width Half Maximum (FWHM) of a Proton Pencil Beam from a 2D dimensional ion chamber array (IBA Matrixx) with limited spatial resolution ( 7.6 mm inter chamber distance). The algorithm would allow beam FWHM measurements to be taken during daily QA without an appreciable time increase.

Methods: Combinations of 147 MeV single spot beams were delivered onto an IBA Matrixx and concurrently on EBT3 films for a standard. Data were collected around the Bragg Peak region and evaluated by a custom MATLAB script based on our algorithm using a least squared analysis. A set of artificial data, modified with random noise, was also processed to test for robustness.

Results: The Matlab script processed Matixx data shows acceptable agreement (within 5%) with film measurements with no single measurement differing by more than 1.8 mm. In cases where the spots show some degree of asymmetry, the algorithm is able to resolve the differences. The algorithm was able to process artificial data with noise up to 15% of the maximum value. Time assays of each measurement took less than 3 minutes to perform, indicating that such measurements may be efficiently added to daily QA treatment.

Conclusion: The developed algorithm can be implemented in daily QA program for Proton Pencil Beam scanning beams (PBS) with Matrixx to extract spot size and position information. The developed algorithm may be extended to small field sizes in photon clinic.