Clinical Benefit for Breast and Chest Wall Setup in Using a Breast Board


S Li

S Li*, C Miyamoto , D Serratore , Q Liang , E Dziemianowicz , Temple University Hospital, Philadelphia, PA

Presentations

SU-F-T-92 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:To validate benefit of using a breast board (BB) by analyzing the geometry and dosimetry changes of the regions of interest (ROIs) between CT scans with and without BB.

Methods:Seven patients, two chest walls (CW) and five breasts, use BB at CT simulation and no BB at diagnostic CT were included. By using deformable image registration software (Velocity AI), diagnostic CT and planning CT were rigidly co-registered according to the thoracic cage at the target. The heart and the target were then deformedly matched and the contours of the planned ROIs were transferred to the diagnostic CT. Which were brought back to the planning CT data set though the initial rigid co-registration in order to keep the deformed ROIs redefined in the diagnostic CT. Anatomic shifts and volume changes of a ROI beyond the rigid translation were recorded and dosimetry changes to ROIs were compared with recalculated DVHs.

Results:Patient setup without the BB had small but systematic heart shifts superiorly by ~5 mm. Torso rotations in two cases moved the heart in opposite directions by ~10 mm. The breast target volume, shape, and locations were significantly changed with arm extension over the head but not in cases with the arm extended laterally. Breast setup without BB could increase the mean dose to the heart and the maximal dose to the anterior ventricle wall by 1.1 and 6.7 Gy, respectively.

Conclusion:A method for evaluation of breast setup technique is introduced and applied for patients. Results of systematic heart displacement without using the BB and the potential increase of heart doses encourage us to further investigate the current trend of not using a BB for easy setup and CT scans. Using a BB would likely increase patient sag during prolonged IMRT and real-time patient position monitoring is clinically desired.