Implementation of a Patient Specific QA Method Using EBT-XD for CyberKnife SRS/SBRT Plans

no image available
K Zerouali

K Zerouali*, J Aubry , R Doucet , Centre hospitalier de l'Universite de Montreal, Montreal, Quebec

Presentations

SU-F-T-569 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:To implement the new EBT-XD Gafchromic films for accurate dosimetric and geometric validation of stereotactic radiosurgery (SRS) and stereotactic body radiation therapy (SBRT) CyberKnife (CK) patient specific QA.

Methods:Film calibration was performed using a triple-channel film analysis on an Epson 10000XL scanner. Calibration films were irradiated using a Varian Clinac 21EX flattened beam (0 to 20 Gy), to ensure sufficient dose homogeneity. Films were scanned to a resolution of 0.3 mm, 24 hours post irradiation following a well-defined protocol. A set of 12 QA was performed for several types of CK plans: trigeminal neuralgia, brain metastasis, prostate and lung tumors. A custom made insert for the CK head phantom has been manufactured to yield an accurate measured to calculated dose registration. When the high dose region was large enough, absolute dose was also measured with an ionization chamber. Dose calculation is performed using MultiPlan Ray-tracing algorithm for all cases since the phantom is mostly made from near water-equivalent plastic.

Results:Good agreement (<2%) was found between the dose to the chamber and the film, when a chamber measurement was possible The average dose difference and standard deviations between film measurements and TPS calculations were respectively 1.75% and 3%. The geometric accuracy has been estimated to be <1 mm, combining robot positioning uncertainty and film registration to calculated dose.

Conclusion:Patient specific QA measurements using EBT-XD films yielded a full 2D dose plane with high spatial resolution and acceptable dose accuracy. This method is particularly promising for trigeminal neuralgia plan QA, where the positioning of the spatial dose distribution is equally or more important than the absolute delivered dose to achieve clinical goals.