Program Information
Increased Skin Dose in Supine Craniospinal Irradiation Due to Carbon Fiber Couch and Vacuum Bag Immobilization Device
D Robertson*, Z Zhao , X Wang , J Yang , UT MD Anderson Cancer Center, Houston, TX
Presentations
SU-F-T-515 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall
Purpose: To measure the surface dose for supine craniospinal irradiation employing posterior beams, treating through an imaging couch and BlueBag immobilization device.
Methods: The percentage depth dose (PDD) in the buildup region of a clinical 6 MV photon beam was measured using an Advanced Markus parallel plate ionization chamber in a solid water phantom. The PDD from a 10x10 cm² anterior beam was measured at 100 cm SSD, simulating a traditional prone craniospinal technique. The measurements were compared to commissioning and treatment planning system data. The PDD was also measured in a posterior setup with the phantom surface laying directly on the Brainlab carbon fiber imaging couch, with the phantom surface 100 cm from the source, simulating a supine craniospinal setup. The posterior measurements were repeated with a BlueBag vacuum immobilization device between the couch and phantom, with thicknesses of 1.7 cm and 5 cm. The PDD from a 10x10 cm² field and a typical 6x30 cm² craniospinal field were also compared. The PDDs were normalized at 5 cm to reflect typical craniospinal prescription dose normalization.
Results: The measured PDD curve from the anterior setup agreed well with commissioning and treatment planning data, with surface doses of 19.9%, 28.8% and 27.7%, respectively. The surface doses of the 10x10 cm² and 6x30 cm² fields delivered through the imaging couch were both 122.4%. The supine setup yielded surface doses of 122.4%, 121.6%, and 119.6% for the couch only, 1.7 cm bag, and 5 cm bag setups, respectively.
Conclusion: Delivering craniospinal irradiation through a carbon fiber couch removes the majority of skin sparing. The addition of a vacuum bag immobilization device restores some skin sparing, but the magnitude of this effect is negligible.