Dose Effects of a 1.5T Magnetic Field On Air-Tissue and Lung-Tissue Interfaces in MRI-Guided Radiotherapy
Xinfeng Chen*, Phillip Prior , Guangpei Chen , Christopher Schultz , X. Allen Li , Department of Radiation Oncology, Medical College of Wisconsin, Milwaukee, WI
Presentations
SU-G-JeP2-5 (Sunday, July 31, 2016) 4:30 PM - 5:00 PM Room: ePoster Theater
Purpose:
The purpose of the study is to investigate the dose effects of electron-return-effect (ERE) at air-tissue and lung-tissue interfaces under a 1.5T transverse-magnetic-field (TMF).
Methods:
IMRT and VMAT plans for representative pancreas, lung, breast and head & neck (H&N) cases were generated following clinical dose volume (DV) criteria. The air-cavity walls, as well as the lung wall, were delineated to examine the ERE. In each case, the original plan generated without TMF is compared with the reconstructed plan (generated by recalculating the original plan with the presence of TMF) and the optimized plan (generated by a full optimization with TMF), using a variety of DV parameters, including V100%, D95% and dose heterogeneity index for PTV, Dmax, and D1cc for OARs (organs at risk) and tissue interface.
Results:
The dose recalculation under TMF showed the presence of the 1.5 T TMF can slightly reduce V100% and D95% for PTV, with the differences being less than 4% for all but lung case studied. The TMF results in considerable increases in Dmax and D1cc on the skin in all cases, mostly between 10-35%. The changes in Dmax and D1cc on air cavity walls are dependent upon site, geometry, and size, with changes ranging up to 15%. In general, the VMAT plans lead to much smaller dose effects from ERE compared to fixed-beam IMRT. When the TMF is considered in the plan optimization, the dose effects of the TMF at tissue interfaces are significantly reduced in most cases.
Conclusion:
The doses on tissue interfaces can be significantly changed by the presence of a 1.5T TMF during MR-guided RT when the TMF is not included in plan optimization. These changes can be substantially reduced or even removed during VMAT/IMRT optimization that specifically considers the TMF, without deteriorating overall plan quality.