Treatment of Gastric MALT Lymphoma Utilizing a Magnetic Resonance Image-Guided Radiation Therapy (MR-IGRT) System: Evaluation of Gating Feasibility
T Mazur*, H Gach , A Chundury , B Fischer-Valuck , J Huang , M Thomas , O Green , Washington University School of Medicine, St. Louis, MO
Presentations
TU-AB-BRA-10 (Tuesday, August 2, 2016) 7:30 AM - 9:30 AM Room: Ballroom A
Purpose: To evaluate the feasibility of real-time, real-anatomy tracking and gating for gastric lymphoma patients treated with magnetic resonance image-guided radiation therapy (MR-IGRT)
Methods: Over the last 2 years, 8 patients with gastric lymphoma were treated with 0.3-T, Co-60 MR-IGRT. Post-treatment analysis of real-time cine imaging in the sagittal plane during each patient’s treatment revealed significant motion of the stomach. While this motion was accounted for with generous PTV margins, the system’s capability for real-time, real-anatomy tracking could be used to reduce treatment margins by gating. However, analysis was needed for the feasibility of gating using only the single available sagittal imaging plane. While any plane may be chosen, if the stomach moves differently where it is not being observed, there may potentially be a mistreatment. To that end, imaging with healthy volunteers was done to ascertain stomach motion over 2-4 min by analyzing multiple parallel sagittal and coronal planes 0.75 cm apart. The stomach was contoured on every slice, and the mean displacement between pairs of contour centroids was used to determine the amount of overall motion.
Results: The mean displacement of the centroid in the image plane was 4.3 ± 0.7 mm. The greatest observed motion was more medial with respect to the patient, and less motion laterally, which implies that gating on a plane located closer to MRI isocenter will provide the more conservative scenario as it will turn the radiation delivery off when the stomach is observed to move outside a predetermined boundary.
Conclusion: The stomach was observed to move relatively uniformly throughout, with maximum extent of motion closer to where most MRI systems have the best spatial integrity (near isocenter). Analysis of possible PTV margins from the healthy volunteer study (coupled with previous patient data on interfraction volumetric stomach deformation) is pending.