2020 Joint AAPM | COMP Virtual Meeting
Back to session list

Session Title: Practical Integration of MRI in Clinical Radiation Oncology
Question 1: 1. Commissioning of an MR Linac should:
Reference:Jackson, E. et al. “Acceptance Testing and Quality Assurance Procedures for Magnetic Resonance Imaging Facilities : Report of MR Subcommittee Task Group 1,” 2010, AAPM Report 100, https://www.aapm.org/pubs/reports/RPT_100.pdf Klein, E. et al. “Task Group 142 report: Quality assurance of medical accelerators,” 2009, Medical Physics, 36 (9), 4197-4212 Huq, S. et al. “The report of Task Group 100 of the AAPM: Application of risk analysis methods to radiation therapy quality management,” 2016, Medical Physics, 43 (7), 4209-4262
Choice A:Follow the recommendations of the task group report on MR-Linac Commissioning published in 2019
Choice B:Be performed according to the guidelines in AAPM TG-142, the same as would be done for a conventional linac
Choice C:Include all elements from both AAPM TG-142 (for the linear acceleratory subsystem) and all elements from the AAPM MR TG-1 (for the MRI System)
Choice D:Include appropriate elements from AAPM commissioning task groups (TG-142 and MR TG-1) and ACR guidance, as well as additional tests based on TG-100 methodologies that reflect how the machine will be used clinically
Question 2: The magnetic field in an MR-Linac poses challenges for commissioning because:
Reference:Meijsing, I., et al. "Dosimetry for the MRI accelerator: the impact of a magnetic field on the response of a Farmer NE2571 ionization chamber." Physics in Medicine & Biology 54.10 (2009): 2993
Choice A:All commissioning measurements of the linac dosimetry system should be performed with the MRI magnet powered off
Choice B:The magnetic field can impact detector performance, and limits the devices and equipment that can be used for measurements
Choice C:The electron return effect causes ion chambers to be unusable
Choice D:The light field from the linac is the only reliable indicator of field size
Question 3: Which of the following is the largest system-induced source of geometric distortion in MR simulation?
Reference:Wang et al., Geometric distortion in clinical MRI systems Part I: evaluation using a 3D phantom. Magnetic Resonance Imaging 22(9) 2004:1211-1221
Choice A:B0 field inhomogeneity
Choice B:Chemical Shift
Choice C:Gradient non-linearity
Choice D:Susceptibility artifacts
Question 4: Which of the following is not a major difference between MR simulation in RT and diagnostic MRI applications
Reference:Paulson et al.,Comprehensive MRI simulation methodology for external beam radiation therapy planning. Medical Physics 42(1) 2015:28-39
Choice A:Immobilization devices
Choice B:RF receiving coil configurations
Choice C:Field of view
Choice D:Magnet Field strength
Choice E:Readout bandwidth
Choice F:Flat tabletop
Question 5: Which of the following is true? In comparison to a prostate volume drawn on MR, the prostate delineated on TRUS and CT are:
Reference:Smith et al, IJROBP 64 (2007) 1238-1247
Choice A:Larger by 16% for TRUS and larger by 16% for CT
Choice B:Smaller by 10% for TRUS and larger by 16% for CT
Choice C:Larger by 16% for TRUS and smaller by 10% for CT
Choice D:Smaller by 10% for TRUS and smaller by 10% for CT
Question 6: Which of the following are commissioning tests that should be conducted when commissioning an applicator for MR based brachytherapy?
Reference:Hellebust et al. Radiotherapy and Oncology 96.2 (2010): 153-160.
Choice A:Identification of the Index position of the most distal dwell position
Choice B:Measurement of the distance of the most distal dwell to outer surface of applicator
Choice C:Spot checks of source path for curved applicators
Choice D:Evaluation of the Imaging artefacts introduced by metallic applicators
Choice E:All of the above
Back to session list