2021 AAPM Virtual Summer School
Back to session list

Session Title: Session 1.1: Overview and Basics
Question 1: What is the advantage of a clinic running their own MR scanner?
Reference:Karlsson et al, Dedicated Magnetic Resonance Imaging in the Radiotherapy Clinic, Int. J. Radiation Oncology Biol. Phys., Vol. 74, No. 2, pp. 644–651, 2009.None, it is always better to use an MR housed in a radiology department
Choice A:Reduced chance of patient-induced distortions
Choice B:The direct transfer of knowledge from CT simulation to MR simulation, eliminating the requirement for new training of simulation therapists
Choice C:The ability to customize the scanner for radiation therapy specific needs such as a flat couch and non-invasive coils
Choice D:None, it is always better to use an MR housed in a radiology department
Question 2: The registration errors associated with fusing MR to CT images:
Reference:Reference: Karlsson et al, Dedicated Magnetic Resonance Imaging in the Radiotherapy Clinic, Int. J. Radiation Oncology Biol. Phys., Vol. 74, No. 2, pp. 644–651, 2009.
Choice A:Are eliminated when the department hosts its own MR scanner
Choice B:Are eliminated with MR-only simulations
Choice C:Are never eliminated since CT is always used in radiation therapy treatment planning
Choice D:Are eliminated by using a 1.5T MR scanner.
Question 3: Radiation Therapy with integrated MR and linear accelerator systems
Reference:Reference: Practical Safety Considerations for Integration of Magnetic Resonance Imaging in Radiation Therapy, Practical Radiation Oncology® (2020) 10, 443-453.
Choice A:Cannot be used for high accuracy treatments because they do not include on-board CT scanners
Choice B:Compromises image quality in favor of more rapid image acquisition (relative to on-board CT systems).
Choice C:Does not need to concern itself with the magnetic field affecting the radiation dose because the radiation dose is delivered by x rays
Choice D:Requires that a clinic integrates and uses MR safety training and awareness
Question 4: The integrated MR systems in MRgRT
Reference:Raaijmakers et al, Integrating a MRI scanner with a 6 MV radiotherapy accelerator: dose increase at tissue-air interfaces in a lateral magnetic field due to returning electrons, PHYSICS IN MEDICINE AND BIOLOGY, Volume: 50 Issue: 7 Pages: 1363-1376
Choice A:Provide radiation doses that are perturbed more by patient density heterogeneities than doses delivered using conventional radiation therapy machines
Choice B:Cannot be operated at the same time as the radiation delivery systems due to the Lorenz force on the linear accelerator
Choice C:Do not need magnetic field shielding to isolate the linear accelerator from the main magnetic field
Choice D:Operate on different physical principles than do conventional MR scanners
Question 5: The Bloch equation dictates
Reference:Reference: F. Bloch, "Nuclear Induction", Physical Review 70, 4604–73 (1946)
Choice A:the motion of the bulk magnetization in the B0 magnetic field
Choice B:how MR images are reconstructed from the k-space data
Choice C:how the concomitant field is generated by the gradient fields
Choice D:the amount of eddy currents generated when running an MRI pulse sequence
Question 6: A gradient field in the X direction is NOT
Reference:Magnetic Resonance Imaging: Physical Principles and Sequence Design, by Robert Brown, PhD et al., Wiley Blackwell, ISBN-13: 978-0471720850, ISBN-10: 0471720852
Choice A:a spatially varying magnetic field
Choice B:a magnetic field that is aligned along the X direction
Choice C:a necessary step in spatial encoding in the X direction in MRI
Choice D:a source of the noise generated by the MRI system when running a pulse sequence
Question 7: emical shift could cause
Reference:Handbook of MRI pulse sequence. Bernstein et al., Elsevier Academic Press, ISBN-13: 978-0-12-092861-3
Choice A:distortion in the reconstructed MR images
Choice B:spatial mis-encoding in MRI
Choice C:signal cancellation
Choice D:all of the above
Question 8: Magnetic field homogeneity over a diameter of spherical volume
Reference:Gach HM, et al, “B0 field homogeneity recommendations, specifications, and measurement units for MRI in radiation therapy”, Med Phys. 2020 Sep;47(9):4101-4114. doi: 10.1002/mp.14306. Epub 2020 Jul 19
Choice A:Decreases as distance from isocenter when measured as root-mean square variance
Choice B:Decreases as distance from isocenter when measured as peak-to-peak change
Choice C:Increases as distance from isocenter
Choice D:Increases with smaller bore size
Question 9: The ‘gradient’ in the x-direction (Gx) is a representation of:
Reference:“The Essential Physics of Medical Imaging”; Bushberg JT, et al, 3rd Edition, Lippincott Williams & Wilkins, 2012
Choice A:Changes in the x-component of the magnetic field in the x-direction
Choice B:Changes in the z-component of the magnetic field in the x-direction
Choice C:Changes in the absolute value of the magnetic field in the x-direction
Choice D:Uncertainty of the magnetic field in the x-direction
Question 10: The spatial heterogeneity in MR phase-array receive coils can be used to encode spatial information. The general term for this form of image acceleration is called:
Reference:“RF coils: A practical guide for nonphysicists”; Bruber B, et al, J Magn Reson Imaging . 2018 Jun 13;48(3):590-604. doi: 10.1002/jmri.26187.
Choice A:Parallel imaging
Choice B:Compressed sensing
Choice C:Parallel transmit
Choice D:Parallel transmit
Back to session list