2021 AAPM Virtual Summer School
Back to session list

Session Title: Session 3.2: MRgRT Commissioning and QA
Question 1: MR to MV transform test measures the alignment between the MR origin and the MV isocenter. Which statement is true about the MR to MV transform?
Reference:Machine QA for the Elekta Unity system: A Report from the Elekta MR-Linac consortium, accepted by Medical Physics (2021)
Choice A:The MR origin overlaps with MV isocenter
Choice B:The MR to MV transform values are very small and it can be ignored
Choice C:The MR to MV transform values are fixed, no need to worry about it
Choice D:The MR to MV transform is measured on machine installation, applied to the treatment planning system and verified periodically
Question 2: The transverse magnetic field will induce which of the following?
Reference:: Relative dosimetry with an MR-linac: Response of ion chambers, diamond, and diode detectors for off-axis, depth dose, and output factor measurements, Med. Phys. 45 (2), (2018), 884-897
Choice A:A lateral shift in the crossline beam profile (perpendicular to the magnetic field)
Choice B:A lateral shift in the inline beam profile (parallel to the magnetic field)
Choice C:No lateral shift in either inline or crossline beam profile
Choice D:lateral shift in both inline and crossline beam profiles
Question 3: Which of the following statements about the MRI gradient fields is false?
Reference:Task Group 284 Report: Magnetic Resonance Imaging Simulation in Radiotherapy: Considerations for Clinical Implementation, Optimization, and Quality Assurance, accepted by Medical Physics (2021)
Choice A:The gradient field encodes spatial information into MR signal
Choice B:Accuracy of spatial mapping relies on the linearity of the gradient system
Choice C:Imperfections in gradient linearity can cause geometric distortion
Choice D:Imperfections in gradient linearity can reduce SNR
Question 4: Regarding treating a patient with a pacemaker on an 0.35T MRIgRT system, which of the following is true?
Reference:Wilkoff, Bruce L., et al. "Magnetic resonance imaging in patients with a pacemaker system designed for the magnetic resonance environment." Heart rhythm 8.1 (2011): 65-73.
Choice A:MRI scanning is never allowed, a pacemaker is an absolute contraindication to MRI
Choice B:If the pacemaker is FDA-cleared for 1.5T only, the manufacturer will support scanning at 0.35T because the lower field is less risky
Choice C:If the patient is not pacemaker dependent, no additional actions are needed
Choice D:Factors potentially affecting the safety of scanning include the field strength, slew rate, SAR, and location of the pacemaker with respect to the RF transmit coil isocenter
Question 5: Issues identified in commissioning a low-field MRI system include which of the following?
Reference:Michael Gach, H., et al. "MRI quality control for low-field MR-IGRT systems: Lessons learned." Journal of applied clinical medical physics 20.10 (2019): 53-66.
Choice A:Gantry-dependent field inhomogeneity changes
Choice B:Image noise due to poorly shielded external RF sources
Choice C:Image noise due to damaged RF coil components
Choice D:All of the above
Question 6: Imaging isocenter variation of an 0.35T MRIgRT system has been observed to vary by up to how many mm with radiotherapy gantry rotation?
Reference:Kim, Taeho, et al. "Characterizing MR imaging isocenter variation in MRgRT." Biomedical Physics & Engineering Express 6.3 (2020): 035009.
Choice A:0.18 mm
Choice B:0.98 mm
Choice C:1.80 mm
Choice D:2.80 mm
Question 7: Which of the following is false regarding reference dosimetry ion chamber correction factors in an 0.35T external magnetic field?
Reference:Krauss, Achim, Claudia Katharina Spindeldreier, and Sebastian Klüter. "Direct determination of for cylindrical ionization chambers in a 6 MV 0.35 T MR-linac." Physics in Medicine & Biology 65.23 (2020): 235049.
Choice A:Larger chambers are expected to exhibit larger correction factors
Choice B:Expected to be smaller than correction factors in a 1.5T field
Choice C:At 1.5T, correction factors are small enough to be safety neglected in all situations
Choice D:Correction factors depend on chamber orientation with respect to magnetic field lines
Question 8: During daily QA, MRI safety quality checks must be performed. Which of the following is recommended to be readily available as per ACR MR safety guidelines?
Reference:ACR Committee on MR safety. ACR Manual on MR safety . Americal College of Radiology. 1891 Preston White Drive, Reston, VA 20191
Choice A:A ferromagnetic detector
Choice B:The presence of hearing protection device
Choice C:The presence of MRI screening forms
Choice D:All of the above
Question 9: The evaluation of spatial accuracy and geometric distortion should be performed on a monthly basis using a large phantom (>30 cm); what is the tolerance?
Reference:Glide-Hurst C., Paulson E., McGee K., Tyagi N., Hu Y., Balter J., Bayouth J., Task Group 284 Report: Magnetic Resonance Imaging Simulation in Radiotherapy: Considerations for Clinical Implementation, Optimization, and Quality Assurance
Choice A:1 mm across the entire FOV
Choice B:2 mm across the entire FOV
Choice C:1 mm for distances less than 10 cm and 2 mm for distances up to 25 cm from isocenter
Choice D:2 mm for distances less than 15 cm and 3 mm for distances up to 35 cm from isocenter
Question 10: AAPM TG 142 provides guidance on quality assurance for linear accelerators, including tolerances for CBCT imaging and radiation coincidence. If similar recommendations are to be adapted for MRI linac, the MR-RT isocenter coincidence tolerance should be which of the following?
Reference:Reference: Klein EE, Hanley J, Bayouth J, et al. Task Group 142 Report: quality assurance of medical accelerators. Med Phys. 2009; 36: 4197– 4212.
Choice A:0.2 mm or less
Choice B:1 mm or less
Choice C:2 mm or less for non-SBRT, 1 mm or less for SBRT-capable machines
Choice D:3 mm or less for non-SBRT, 2 mm or less for SBRT-capable machines
Choice E:None of the above
Back to session list