Encrypted login | home

Program Information

Influence of Scanner Output Measurement Technique On KERMA Ratios in CT


K Ogden

K Ogden*, M Roskopf, E Scalzetti, SUNY Upstate Medical Univ, Syracuse, NY

SU-E-I-49 Sunday 3:00:00 PM - 6:00:00 PM Room: Exhibit Hall

Purpose: KERMA ratios (RK) are defined as the ratio of KERMA measured at a specific phantom location (K) to in-air isocenter CT scanner output (KCT). In this work we investigate the impact of measurement methodology on KCT values.
Methods: OSL dosimeter chips were used to measure KCT for a GE VCT scanner (GE Medical Systems, Waukesha WI), using the 40 mm nominal beam width. Methods included a single point measurement at the center of the beam (1 tube rotation), and extended z-axis measurements using multiple adjacent OSL's (7.5 cm extent), with single tube rotation, multiple contiguous axial scans, and helical scans (pitch of 1.375). Measurements were made in air and on the scan table at 80 and 120 kV.
Results: Averaged single point measurements were consistent, with a mean coefficient of variation of 2.5%. For extended measurements with a single tube rotation, the mean value was equivalent to the single point measurements. For multiple contiguous axial scans, the in-air KCT values were higher than the single rotation mean value and single point measurements by 13% and 10.3% at 120 and 80 kV, respectively, and for the on-table measurements the values were 14.9% and 8.1% higher at 120 and 80 kV, respectively. The increase is due to beam overlap caused by z-axis over-beaming. Extended measurements using helical scanning were equivalent to the multiple rotation axial measurements when corrected for the helical pitch. For all methodologies, the in-air values exceeded the on-table measurements by an average of 23% and 19.4% at 80 and 120 kV, respectively.
Conclusions: Scanner KCT values must be measured to allow organ dose estimation using published RK values. It is imperative that the KCT measurement methodology is the same as for the published values, or large errors may be introduced into the resulting organ dose estimates.


Contact Email