Encrypted login | home

Program Information

Electron Beam Profile Differences Between Elekta MLCi2 and Elekta Agility Treatment Heads


C Wu

C Wu1*, C Hatcher2 , (1) Sutter Medical Foundation, Roseville, CA 95661, (2) Sutter Shared Services S3, Sacramento, California

Presentations

SU-E-T-149 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose: To report and investigate observed differences in electron beam profiles at various energies/applicators between Elekta MLCi2 and Agility treatment head on Elekta Infinity LINAC
Methods: When we upgraded from MLCi2 to Agility on one of our Elekta Infinity LINAC's, electron beam PDDs and profiles were acquired for comparison purpose. All clinical electron energies (6/9/12/15/12/18 MeV) and electron applicators (6/10/14/20/25 square) were included in measurement. PDDs were acquired at 100 SSD in water (PTW MP3 water tank) with a plane-parallel ion chamber (PTW Roos). X and Y Profiles were acquired using IC Profiler (Sun Nuclear Corp.) at 1cm and maximum PDD depths (water equivalent).
Results: All PDD curves match very well between MLCi2 and Agility treatment head. However, some significant differences on electron profiles were found. On Agility, even after increasing the default auto-tracking offset values for backup diaphragms in Y and MLC in X by 2.8 cm (the maximum allowed change is 3.0 cm), electron profiles still have rounder shoulders comparing to corresponding MLCi2 profiles. This difference is significantly more pronounced at larger applicators (20 and 25 square), for all electron energies.
Conclusion: The significant design change between MLCi2 and Agility beam limiting device seems to affect exit electron beam profiles. In IEC1217 X direction, the main change on Agility is the removal of the original MLCi2 X backup diaphragms and replacing it with MLC leaves; In Y direction, the main change is the radius and materials on Y backup diaphragms.



Contact Email: