Encrypted login | home

Program Information

Optimizing the MLC Model Parameters for IMRT in the RayStation Treatment Planning System

no image available
S Chen

S Chen*, B Yi , H Xu , X Yang , K Prado , W D'Souza , University of Maryland School of Medicine, Baltimore, MD

Presentations

SU-E-T-583 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose: To optimize the MLC model parameters for IMRT in the RayStation v.4.0 planning system and for a Varian C-series Linac with a 120-leaf Millennium MLC.

Methods: The RayStation treatment planning system models rounded leaf-end MLC with the following parameters: average transmission, leaf-tip width, tongue-and-groove, and position offset. The position offset was provided by Varian. The leaf-tip width was iteratively evaluated by comparing computed and measured transverse dose profiles of MLC-defined fields at dmax in water. The profile comparison was also used to verify the MLC position offset. The transmission factor and leaf tongue width were derived iteratively by optimizing five clinical patient IMRT QA results: brain, lung, pancreas, head-and-neck (HN), and prostate. The HN and prostate cases involved splitting fields. Verifications were performed with Mapcheck2 measurements and Monte Carlo calculations. Finally, the MLC model was validated using five test IMRT cases from the AAPM TG119 report. Absolute gamma analyses (3mm/3% and 2mm/2%) were applied. In addition, computed output factors for MLC-defined small fields (2X2, 3X3, 4X4, 6X6cm) of both 6MV and 18MV were compared to those measured by the Radiological Physics Center (RPC).

Results: Both 6MV and 18MV models were determined to have the same MLC parameters: 2.5% transmission, tongue-and-groove 0.05cm, and leaf-tip 0.3cm. IMRT QA analysis for five cases in TG119 resulted in a 100% passing rate with 3mm/3% gamma analysis for 6MV, and >97.5% for 18MV. With 2mm/2% gamma analysis, the passing rate was >94.6% for 6MV and >90.9% for 18MV. The difference between computed output factors in RayStation and RPC measurements was less than 2% for all MLC-defined fields, which meets the RPC’s acceptance criterion.

Conclusion: The rounded leaf-end MLC model in RayStation 4.0 planning system was verified and IMRT commissioning was clinically acceptable. The IMRT commissioning was well validated using guidance from the AAPMTG119 protocol.


Contact Email: