IBA ProteusOne Compact Proton Therapy System Radiation Survey Results


J Zhang

J Zhang*, J Syh , J Syh , M White , B Patel , X Song , H Wu , Willis-Knighton Cancer Center, Shreveport, LA

Presentations

SU-E-T-200 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose:
This study summarizes the results of an initial radiation survey of the Willis-Knighton Cancer Center in Shreveport, Louisiana. The facility houses an IBA ProteusOne compact single room proton therapy unit coupled with a C230 cyclotron that operates at a maximum energy of 230 MeV.

Methods:
A calibrated survey meter was used for the photon measurements to obtain reliable results. A neutron detector was used as the measuring instrument for neutrons. The locations of the survey and measurements were planned carefully in order to get a proper evaluation of the facility shielding configuration. The walls, ceiling, vault entrance, and the adjacent environment were each surveyed with suitable measurement instruments. A total of 22 locations were chosen for radiation survey. Dose equivalent values were calculated for both the photon and the neutron radiation using measured data.

Results:
All measured dose values are presented in millisievert per year. The highest dose measured at the vault entrance is 0.34 mSv/year. A dedicated shielding door was not present at the time of the measurement. The vault entrance area is considered as a controlled area. The shielding design goals are not to exceed 5 mSv/year for the controlled area and 1 mSv/year the uncontrolled area. The total combined neutron and photon dose equivalent values were found to be compliant with the Louisiana Department of Environmental Quality radiation protection regulatory codes.

Conclusion:
In our efforts to evaluate the radiation levels at the Willis-Knighton Cancer Center proton treatment facility, we have found that all the measured values of the radiation shielding are below the critical radiation limits per year. Since the total dose measured at the vault entrance is below the shielding design goal, a shielding door is not required at this proton treatment vault.