Encrypted login | home

Program Information

Dose Delivery Accuracy in Breast Field Junction for Free Breath and Deep Inspiration Breath Hold Techniques

no image available
D Epstein

D Epstein*, E Shekel , D Levin , Assuta Medical Centers, Tel Aviv

Presentations

SU-E-T-426 Sunday 3:00PM - 6:00PM Room: Exhibit Hall

Purpose: The purpose of this work was to verify the accuracy of the dose distribution along the field junction in a half beam irradiation technique for breast cancer patients receiving radiation to the breast or chest wall (CW) and the supraclavicular LN region for both free breathing and deep inspiration breath hold (DIBH) technique.

Methods: We performed in vivo measurements for nine breast cancer patients receiving radiation to the breast/CW and to the supraclavicular LN region. Six patients were treated to the left breast/CW using DIBH technique and three patients were treated to the right breast/CW in free breath. We used five microMOSFET dosimeters: three located along the field junction, one located 1 cm above the junction and the fifth microMOSFET located 1 cm below the junction. We performed consecutive measurements over several days for each patient and compared the measurements to the TPS calculation (Eclipse, Varianâ„¢).

Results: The calculated and measured doses along the junction were 0.97±0.08 Gy and 1.02±0.14 Gy, respectively. Above the junction calculated and measured doses were 0.91±0.08 Gy and 0.98±0.09 Gy respectively, and below the junction calculated and measured doses were 1.70±0.15 Gy and 1.61±0.09 Gy, respectively. All differences were not statistically significant. When comparing calculated and measured doses for DIBH patients only, there was still no statistically significant difference between values for all dosimeter locations. Analysis was done using the Mann-Whitney Rank-Sum Test.

Conclusion: We found excellent correlation between calculated doses from the TPS and measured skin doses at the junction of several half beam fields. Even for the DIBH technique, where there is more potential for variance due to depth of breath, there is no over or underdose along the field junction. This correlation validates the TPS, as well an accurate, reproducible patient setup.


Contact Email: