Monitor Unit Optimization in Stereotactic Body Radiation Therapy for Stage I Lung Cancer

no image available
B Huang

B-T Huang, J-Y Lu* Cancer Hospital of Shantou University Medical College, Shantou, China

Presentations

SU-E-T-551 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:
The study aims to reduce the monitor units (MUs) in the stereotactic body radiation therapy (SBRT) treatment for lung cancer by adjusting the optimizing parameters.

Methods:
Fourteen patients suffered from stage I Non-Small Cell Lung Cancer (NSCLC) were enrolled. Three groups of parameters were adjusted to investigate their effects on MU numbers and organs at risk (OARs) sparing: (1) the upper objective of planning target volume (UOPTV); (2) strength setting in the MU constraining objective; (3) max MU setting in the MU constraining objective.

Results:
We found that the parameters in the optimizer influenced the MU numbers in a priority, strength and max MU dependent manner. MU numbers showed a decreasing trend with the UOPTV increasing. MU numbers with low, medium and high priority for the UOPTV were 428±54, 312±48 and 258±31 MU/Gy, respectively. High priority for UOPTV also spared the heart, cord and lung while maintaining comparable PTV coverage than the low and medium priority group. It was observed that MU numbers tended to decrease with the strength increasing and max MU setting decreasing. With maximum strength, the MU numbers reached its minimum while maintaining comparable or improved dose to the normal tissues. It was also found that the MU numbers continued to decline at 85% and 75% max MU setting but no longer to decrease at 50% and 25%. Combined with high priority for UOPTV and MU constraining objectives, the MU numbers can be decreased as low as 223±26 MU/Gy.

Conclusion:
The priority of UOPTV, MU constraining objective in the optimizer impact on the MU numbers in SBRT treatment for lung cancer. Giving high priority to the UOPTV, setting the strength to maximum value and the max MU to 50% in the MU objective achieves the lowest MU numbers while maintaining comparable or improved OAR sparing.