Pinnacle-Based Tool to Evaluate the Effect of Prostate Rotation as Determined by Calypso

no image available
R Velasco-Schmitz

S Narayanan*, R Velasco-Schmitz, K Claeys, P Cho , Virginia Mason Cancer Institute, Seattle, WA

Presentations

SU-E-P-12 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: When the Calypso reports a large prostate rotation and one does not have the means to make corrections, it is helpful to quantify its dosimetric effects. We have devised a simple scheme to achieve this in Pinnacle TPS. If it is determined that the PTV margin is inadequate, an alternate larger-margin plan may be adapted for the day.

Methods: A three-step process using the Pinnacle Image fusion module was formulated to analyze the clinical effect of prostate rotation. The process includes (1) translating the image set such that the rotational axis is about the isocenter, (2) performing 3D rotations of the structures of interest, and (3) restoring the coordinates associated with the plan. The rotated structures are imported and overlayed on the original plan for evaluation.
The tool was applied to three prostate cancer patients with relatively large rotations (95 fractions total). Prostate rotations are primarily due to variations in bladder and rectal filling, which may also affect the seminal vesicles. It is difficult to estimate the perturbation of seminal vesicles for a given prostate rotation. Therefore, the Calypso-derived rotation matrix was applied only to CTV and PTV (CTV+5mm).


Results: The rotations were predominantly in the pitch direction with an absolute value ranging from 0-29 degrees (Median = 10°, Average = 14.5°). For various magnitudes of rotation, the CTV V100 decreased from the original 100% to 99.6%/99%/95% for 10°/15°/29° rotations, respectively. Likewise, the PTV V100 diminished from 95% to 92%/89%/82%.

Conclusion: A tool, consisting only of features within a TPS, was used to evaluate the dosimetric effect of prostate rotation. Our study shows that under large uncorrected rotations, the tumor coverage is degraded and may require intervention in the form of plan adaptation. To this end, alternate plans with variable margins can be generated in advance for daily adaption.