Encrypted login | home

Program Information

A Research of PET-CT SUV Range for the Online Dose Verification in Carbon Ion Radiation Therapy

no image available
L Sun

L Sun1*, W Hu2 , M Moyers3 , J Zhao4 , W Hsi5 , (1) Fudan University Shanghai Cancer Center, Shanghai, Shanghai, (2) Fudan University Shanghai Cancer Center, Shanghai, Shanghai, (3) Shanghai Proton and Heavy Ion Center, Colton, CA, (4) Shanghai Proton and Heavy Ion Center, Shanghai, Shanghai, (5) Shanghai Proton and Heavy Ion Center, Shanghai, China,

Presentations

SU-E-J-146 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:Positron-emitting isotope distributions can be used for the image fusion of the carbon ion planning CT and online target verification PET-CT,after radiation in the same decay period,the relationship between the same target volume and the SUV value of different every single fraction dose can be found,then the range of SUV for the radiation target could be decided.So this online range also can provide reference for the correlation and consistency in planning target dose verification and evaluation for the clinical trial.

Methods:The Rando head phantom can be used as real body,the 10cc cube volume target contouring is done,beam ISO Center depth is 7.6cm and the 90 degree fixed carbon ion beams should be delivered in single fraction effective dose of 2.5GyE,5GyE and 8GyE.After irradiation,390 seconds later the 30 minutes PET-CT scanning is performed,parameters are set to 50Kg virtual weight,0.05mCi activity.MIM Maestro is used for the image processing and fusion,five 16mm diameter SUV spheres have been chosen in the different direction in the target.The average SUV in target for different fraction dose can be found by software.

Results:For 10cc volume target,390 seconds decay period,the Single fraction effective dose equal to 2.5Gy,Ethe SUV mean value is 3.42,the relative range is 1.72 to 6.83;Equal to 5GyE,SUV mean value is 9.946,the relative range is 7.016 to 12.54;Equal or above to 8GyE,SUV mean value is 20.496,the relative range is 11.16 to 34.73.

Conclusion:Making an evaluation for accuracy of the dose distribution using the SUV range which is from the planning CT with after treatment online PET-CT fusion for the normal single fraction carbon ion treatment is available.Even to the plan which single fraction dose is above 2GyE,in the condition of other parameters all the same,the SUV range is linearly dependent with single fraction dose,so this method also can be used in the hyper-fraction treatment plan.


Contact Email: