Encrypted login | home

Program Information

Characteristics of X-Rays From ExacTrac and Patient Dose From Imaging Procedures


G Ding

G Ding*, Vanderbilt University, Nashville, TN

Presentations

SU-E-J-23 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:
The purpose of this investigation is 1) provide the beam characteristics of x-rays produced by a Novalis TX ExacTrac system; 2) present a method to commission such beams, 3) present radiation dose to patients resulting from the imaging procedures.

Methods:
The Monte Carlo simulations were used to obtain the characteristics of kV beams and validated by measurements. The calculated beam HVLs, profiles and depth-dose curves were benchmarked against measurements. Twelve different image acquisition protocols were studied. The x-ray tube voltage ranges from 70 - 145 kV and milliampere-second (mAs) ranges from 8 - 80 mAs depending on the selection of Cranium, Head & Neck, Thorax or Abdomen imaging protocols. The beam output of each image acquisition protocol was determined by using an ionization chamber. The air kerma calibration factors of the ion chamber were obtained from an Accredited Dosimetry Calibration Laboratory for specified HVLs.

Results:
The agreements between measured and simulated results were within the uncertainties for HVLs, dose profiles and depth-dose curves. When %dd was normalized at 1 cm depth, its values at depth of 5 cm ranged from 45% to 66% of in water for kV beams range from 70 kVp to 145 kVp respectively. For head images, a typical dose to eye resulting from single projection ranged from 0.008 cGy to 0.025 cGy depending on the cranial image protocols selected. For a single pelvic image the skin dose could reach up to 0.1 cGy from an abdominal protocol.

Conclusion:
Although multiple pairs of x-ray images are commonly acquired during a daily patient treatment, the imaging doses to patients resulting from the sum of these projected x-rays are generally much less than 0.5 cGy. The knowledge obtained from this investigation can be used to estimate the image dose and optimize the used of the system.


Contact Email: