Encrypted login | home

Program Information

A Feasibility Study of Using Eclipse AAA for SRS Treatement

no image available
S Lim

S Lim*, T LoSasso , Memorial Sloan-Kettering Cancer Center, New York, NY

Presentations

SU-E-P-16 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:
To commission Varian Eclipse AAA for SRS treatment and compare the accuracy with Brainlab iPlan system for clinical cases measured with radiochromic film

Methods:
A 6MV AAA clinical model for a Varian TrueBeam STx is used as baseline. The focal spot and field size of the baseline model(BASE) are (1.75,0.75) and 40x40cm² respectively. Maximum field sizes, output factors(St), FWHM focal spot and secondary source sizes are systematically adjusted to obtain an optimized model(OPT) by comparing the calculated PDD’s, profiles, and output factors with measurements taken with a stereotactic diode(SD) and, cc01 and cc04 ion chambers in Blue Phantom. In-phantom dose distributions of clinical SRS fields are calculated using the OPT and the clinical Brainlab iPlan pencil-beam. Within the 90% isodose-line(ROI), the average dose difference between the calculations and radiochromic film measurements are assessed.

Results:
The maximum field, focal spot and secondary source sizes for the OPT are 15x15cm², (0,0), and 32.3mm respectively. The OPT St input at 1.0 and 2.0cm fields are increased by 4.5% and 1.5% from BASE. The calculated output of the BASE and OPT underestimate by 16.1%-3.2% respectively at 0.5x0.5cm² field and 3.1%-0.02% respectively at 1.0x1.0cm² field. The depth doses at 10cm are within 3.5% and 0.4% of measurements for 0.5x0.5 and 1.0x1.0cm². The ROI dose of OPT and iPlan are within 1.6% and 0.6% of film measurements for 3.0cm clinical fields. For 1.0cm fields, the ROI dose of OPT underestimate 0.0-2.0% and iPlan overestimates 1.7-2.9% relative to measurements.

Conclusion:
The small field dose calculation of Eclipse AAA algorithm can be significantly improved by carefully adjusting the input parameters. The larger deviation of the OPT for 0.5x0.5cm² field from measurements can be attributed to the lowest 1.0cm field size input limit of AAA. The OPT compares reasonably well with the iPlan pencil-beam and measurements.


Contact Email: