Geometric Uncertainties in Eye Plaque Dosimetry for a Fully Loaded 16 Mm COMS Plaque


H Morrison

H Morrison*, G Menon , H Jans , M Larocque , R Sloboda , Cross Cancer Institute, Edmonton, AB, and University of Alberta, Edmonton, AB

Presentations

SU-E-T-443 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: To determine the effect of geometric uncertainties in the seed positions in a COMS eye plaque on the central axis (CAX) dose.

Methods: A Silastic insert was placed into a photopolymer 3D printed 16 mm COMS plaque, which was then positioned onto a custom-designed PMMA eye phantom. High resolution 3D images were acquired of the setup using a Siemens Inveon microPET/CT scanner. Images were acquired with the plaque unloaded and loaded with IsoAid I-125 seed shells (lack of silver core to minimize metal artifacts). Seed positions and Silastic thickness beneath each slot were measured. The measured seed coordinates were used to alter the seed positions within a standard 16 mm COMS plaque in Plaque Simulator v5.7.3 software. Doses along the plaque CAX were compared for the original and modified plaque coordinates using 3.5 mCi seeds with treatment times set to deliver 70 Gy to tumour apexes of 3.5, 5, and 10 mm height.

Results: The majority of seeds showed length-wise displacement, and all seeds showed displacement radially outward from the eye center. The average radial displacement was 0.15 mm larger than the expected 1.4 mm offset, approximately half of which was due to increased Silastic thickness beneath each slot. The CAX doses for the modified seed positions were consistently lower for all tumour heights due to geometric displacement of the seeds; dose differences were found to increase to a maximum of 2.6% at a depth of ~10 mm, after which they decreased due to the inverse square dose fall-off minimizing this effect.

Conclusion: This work presents initial results of a broader dosimetric uncertainty evaluation for fully loaded COMS eye plaques and demonstrates the effects of seed positioning uncertainties. The small shifts in seed depths had noticeable effects on the CAX doses indicating the importance of careful Silastic loading.