Encrypted login | home

Program Information

Development of a HDR-BT QA Tool for Verification of Source Position with Oncentra Applicator Modeling

no image available
Y Kumazaki

Y Kumazaki*, K Miyaura , R Hirai , K Miyazawa , S Makino , T Tamaki , N Shikama , S Kato , Saitama Medical University International Medical Center, Hidaka, Saitama

Presentations

SU-E-T-254 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:To develop a High Dose Rate Brachytherapy (HDR-BT) quality assurance (QA) tool for verification of source position with Oncentra applicator modeling, and to report the results of radiation source positions with this tool.

Methods:We developed a HDR-BT QA phantom and automated analysis software for verification of source position with Oncentra applicator modeling for the Fletcher applicator used in the MicroSelectron HDR system. This tool is intended for end-to-end tests that mimic the clinical 3D image-guided brachytherapy (3D-IGBT) workflow. The phantom is a 30x30x3 cm cuboid phantom with radiopaque markers, which are inserted into the phantom to evaluate applicator tips and reference source positions; positions are laterally shifted 10 mm from the applicator axis. The markers are lead-based and scatter radiation to expose the films. Gafchromic RTQA2 films are placed on the applicators. The phantom includes spaces to embed the applicators. The source position is determined as the distance between the exposed source position and center position of two pairs of the first radiopaque markers. We generated a 3D-IGBT plan with applicator modeling. The first source position was 6 mm from the applicator tips, and the second source position was 10 mm from the first source position.

Results: All source positions were consistent with the exposed positions within 1 mm for all Fletcher applicators using in-house software. Moreover, the distance between source positions was in good agreement with the reference distance. Applicator offset, determined as the distance from the applicator tips at the first source position in the treatment planning system, was accurate.

Conclusion: Source position accuracy of applicator modeling used in 3D-IGBT was acceptable. This phantom and software will be useful as a HDR-BT QA tool for verification of source position with Oncentra applicator modeling.




Contact Email: