Encrypted login | home

Program Information

Dosimetric Comparison of Leaf with Or Without Interdigitation in Multiple Brain Metastasis VMAT Treatment Planning

no image available
J Duan

Department of Radiation Oncology, Shandong Tumor Hospital, Jinan, China

Presentations

SU-E-T-306 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: To evaluate the effects of leaf with or without interdigitation in multiple brain metastasis volumetric modulated arc therapy (VMAT) plans.
Methods: Twenty patients with 2 to 6 brain metastases of our hospital were retrospectively studied to be planned with dual arc VMAT using Monaco 3.3 TPS on the Elekta Synergy linear accelerator. The prescription dose of PTV was 60Gy/30 fractions. Two plans with or without leaf interdigitation were designed. The homogeneity index (HI), conformity index (CI), dose volume histograms (DVHs), monitor unit (MU), treatment time (T), the segments, the dose coverage of the target, were all evaluated.
Results: The plans with leaf interdigitation could achieve better CI (p<0.05) than without leaf interdigitation, while no significant difference were found in HI (p> 0.05) and the dose coverage of the target (p> 0.05) .The MU,T, and the segments of the plan with leaf interdigitation were more than the plan without leaf interdigitation (p<0.05). There was no significant difference found in radiation dose of spinal cord, lenses and parotids, while the maximum dose of brain stem of leaf without interdigitation was higher than leaf with interdigitation (p< 0.05). It was worth noting that the areas of low dose regions with leaf interdigitation plan were much less than the without leaf interdigitation plan in the doublication planes (p< 0.05).
Conclusion: This study shows that leaf with interdigitation has some advantages than leaf without interdigitation in multiple brain metastasis VMAT plans although the clinical relevance remains to be proven.



Contact Email: