Characterization of a Novel High Resolution 1D Silicon Monolithic Array for Small Field Commissioning and Quality Assurance

no image available
F Bisello

F Bisello2,3*, J McGlade1 , P Wang1 , J Kralik1 , P Kosterin1 , R Mooij1 , D Menichelli3 , J Celi3 , T Solberg1 , (1) University of Pennsylvenia, Philadelphia, PA,(2) FAU University Erlangen-Nuernberg, Erlangen, Germany (3) IBA Dosimetry, Schwarzenbruck, Germany

Presentations

SU-E-T-163 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:To study the suitability of a novel 1D silicon monolithic array for dosimetry of small radiation fields and for QA of high dose gradient treatment modalities (IMRT and SBRT).

Methods:A 1D array composed of 4 monolithic silicon modules of 64 mm length and 1 mm pixel pitch was developed by IBA Dosimetry. Measurements were carried out for 6MV and 15MV photons on two commercial different linacs (TrueBeam and Clinac iX, Varian Medical Systems, Palo Alto, CA) and for a CyberKnife G4 (Accuray Inc., Sunnyvale, CA). The 1D array was used to measure output factors (OF), profiles and off axis correction factors (OACF) for the Iris CyberKnife variable collimator (5-60 mm). In addition, dose profiles (at the isocenter plane) were measured for multiple IMRT and SBRT treatment plans and compared with those obtained using EDR2radiographic film (Carestream Health, Rochester NY), a commercial 2D diode array and with the dose distribution calculated using a commercial TPS (Eclipse, Varian Medical Systems, Palo Alto, CA).

Results:Due to the small pixel pitch of the detector, IMRT and SBRT plan profiles deviate from film measurements by less than 2%. Similarly, the 1D array exhibits better performance than the 2D diode array due to the larger (7 mm) pitch of that device. Iris collimator OFs measured using the 1D silicon array are in good agreement with the commissioning values obtained using a commercial stereotactic diode as well as with published data. Maximum deviations are < 3% for the smallest field (5 and 7.5mm) and below 1% for all other dimensions.

Conclusion:We have demonstrated good performances of the array for commissioning of small photon fields and in patient QA, compared with diodes and film typically used in these clinical applications. The technology compares favorably with existing commercial solutions

Funding Support, Disclosures, and Conflict of Interest: The presenting author is founded by a Marie Curie Early Initial Training Network Fellowship of the European Communitys Seventh Framework Programme under contract number (PITN-GA-2011-289198-ARDENT). The research activity is hosted by IBA Dosimetry, Gmbh.