Evaluation of MLC Positioning Errors in Dynamic IMRT Treatments by Analyzing Dynalog Files

no image available
J Olasolo

J Olasolo1*, S Pellejero1 , M Gracia1 , N Gallardo1 , M Martin1 , S Lozares1 , F Maneru1 , L Bragado1 , S Miquelez1 , A Rubio1 , (1) Complejo Hospitalario de Navarra, Pamplona, Navarra

Presentations

SU-E-P-36 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:To assess the accuracy of MLC positioning in Varian linear accelerator, in dynamic IMRT technique, from the analysis of dynalog files generated by the MLC controller.

Methods:In Clinac accelerators (pre-TrueBeam technology), control system has an approximately 50ms delay (one control cycle time). Then, the system compares the measured position to the planned position corresponding to the next control cycle. As it has been confirmed by Varian technical support, this effect causes that measured positions appear in dynalogs one cycle out of phase with respect to the planned positions.
Around 9000 dynalogs have been analyzed, coming from the three linear accelerators of our center (one Trilogy and two Clinac 21EX) equipped with a Millennium 120 MLC. In order to compare our results to recent publications, leaf positioning errors (RMS and 95th percentile) are calculated with and without delay effect. Dynalogs have been analyzed using a in-house Matlab software.

Results:The RMS errors were 0.341, 0.339 and 0.348mm for each Linac; being the average error 0.343 mm. The 95th percentiles of the error were 0.617, 0.607 and 0.625; with an average of 0.617mm. A recent multi-institution study carried out by Kerns et al. found a mean leaf RMS error of 0.32mm and a 95th percentile error value of 0.64mm.
Without delay effect, mean leaf RMS errors obtained were 0.040, 0.042 and 0.038mm for each treatment machine; being the average 0.040mm. The 95th percentile error values obtained were 0.057, 0.058 and 0.054 mm, with an average of 0.056mm.

Conclusion:Results obtained for the mean leaf RMS error and the mean 95th percentile were consistent with the multi-institution study. Calculated error statistics with delay effect are significantly larger due to the speed proportional and systematic leaf offset. Consequently it is proposed to correct this effect in dynalogs analysis to determine the MLC performance.