Neutron Dose Survey of a Compact Single Room Proton Machine


Y Chen

Y Chen*, M Prusator , M Islam , D Johnson , S Ahmad , University of Oklahoma Health Sciences Center, Oklahoma City, OK

Presentations

SU-E-T-568 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:To ensure acceptable radiation limits are maintained for those working at and near the machine during its operation, a comprehensive radiation survey was performed prior to the clinical release of Mevion S250 compact proton machine at Stephenson Oklahoma Cancer Center.

Methods:The Mevion S250 proton therapy system consists of the following: a superconducting cyclotron to accelerate the proton particles, a passive double scattering system for beam shaping, and paired orthogonal x-ray imaging systems for patient setup and verification via a 6D robotic couch. All equipment is housed within a single vault of compact design. Two beam delivery applicators are available for patient treatment, offering field sizes of as great as 14 cm and 25 cm in diameter, respectively. Typical clinical dose rates are between 1 and 2 Gy/min with a fixed beam energy of 250 MeV. The large applicator (25 cm in diameter) was used in conjunction with a custom cut brass aperture to create a 20 cm x 20 cm field size at beam isocenter. A 30 cm x 30 cm x 35 cm high density plastic phantom was placed in the beam path to mimic the conditions creating patient scatter. Measurements integrated-ambient-neutron-dose-equivalence were made with a SWENDII detector. Gantry angles of 0, 90 and 180 degrees, with a maximum dose rate of 150 MU/min (for large applicator) and beam configuration of option 1 (range 25 cm and 20 cm modulation), were selected as testing conditions. At each point of interest, the highest reading was recorded at 30 cm from the barrier surface.

Results: The highest neutron dose was estimated to be 0.085 mSv/year at the console area.

Conclusion:All controlled areas are under 5 mSv/year and the uncontrolled areas are under 1 mSv/year. The radiation protection provided by the proton vault is of sufficient quality.


Contact Email: