Performance of MR-Based 3D FXG Dosimetry for Preclinical Irradiation

no image available
M Welch

M Welch1*, D Jaffray1234 , (1) Department of Medical Biophysics, University of Toronto, Toronto, ON,(2) Radiation Medicine Program, Princess Margaret Cancer Centre, Toronto, ON, (3) Department of Radiation Oncology, University of Toronto, Toronto, ON, (4) TECHNA Institute for the Advancement of Technology for Health, Toronto, ON

Presentations

SU-E-T-606 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:
Technological advances have revolutionized preclinical radiation research to enable precise radiation delivery in preclinical models. Kilovoltage x-rays and complex geometries in preclinical radiation studies challenge conventional dosimetry methods. Previously developed gel-based dosimetry provides a viable means of accommodating complex geometries and accurately reporting dose at kV energies. This paper will describe the development and evaluation of gel-based ferrous xylenol-orange (FXG) dosimetry using a 7T preclinical imaging system.

Methods:
To confirm water equivalence, Zeff values were calculated for the FXG material, water and ICRU defined soft tissue. Proton T1 relaxivity response in FXG was measured using a preclinical 7T MR and a small animal irradiator for a dose range of 1-22 Gy. FXG was contained in 50 ml centrifuge tubes and irradiated with a 225 kVp x-ray beam at a nominal dose rate of 2.3 Gy/min. Pre and post irradiation maps of the T1 relaxivity were collected using variable TR spin-echo imaging (TE 6.65 ms; TR 500, 750, 1000, 1500, 2000, 3000 and 5000 ms) with 2 mm thick slices, 0.325 mm/pixel, 3 averages and an acquisition time of 26 minutes. A linear fit to the change in relaxation rate (1/T1) for the delivered doses reported the gel sensitivity in units of ms⁻¹Gy⁻¹. Irradiation and imaging studies were repeated using three batches of gel over 72 hrs.

Results:
FXG has a Zeff of 3.8 for the 225 kVp spectrum used; differing from water and ICRU defined soft tissue by 0.5% and 2.5%, respectively. The average sensitivity for the FXG dosimeter was 31.5 ± 0.7 ms⁻¹Gy⁻¹ (R² = 0.9957) with a y-intercept of -29.4 ± 9.0 ms⁻¹.

Conclusion:
Preliminary results for the FXG dosimeter properties, sensitivity, and dose linearity at preclinical energies is promising. Future work will explore anatomically relevant tissue inclusions to test MR performance.

Funding Support, Disclosures, and Conflict of Interest: Student funding provided by The Terry Fox Foundation Strategic Initiative for Excellence in Radiation Research for the 21st Century at CIHR and the Gifford Ontario Student Opportunity Trust Fund.